Publications
Detailed Information
Rapid Prototyping으로 제작한 3D Medical Model의 오차 측정에 관한 연구 (임상 적용 가능성 및 사례) : Accuracy tests of 3D rapid prototyping(RP) medical models: Its potential and clinical applications
DC Field | Value | Language |
---|---|---|
dc.contributor.author | 최진영 | - |
dc.contributor.author | 최정호 | - |
dc.contributor.author | 김남국 | - |
dc.contributor.author | 이종기 | - |
dc.contributor.author | 김명기 | - |
dc.contributor.author | 김명진 | - |
dc.contributor.author | 김영호 | - |
dc.date.accessioned | 2011-10-19 | - |
dc.date.available | 2011-10-19 | - |
dc.date.issued | 1999 | - |
dc.identifier.citation | 대한구강악안면외과학회지 25(4):295-303,1999 | en |
dc.identifier.issn | 1225-1585 | - |
dc.identifier.uri | https://hdl.handle.net/10371/74300 | - |
dc.description.abstract | Presented in this paper are the experimental results that measure rapid prototyping (RP) errors in 3D medical models. We identified various
factors that can cause dimensional errors when producing RP models, specifically in maxillofacial areas. For the experiment, we used a human dry skull. A number of linear measurements based on landmarks were first obtained on the skull. This was followed by CT scanning, 3D model reconstruction, and RP model fabrication. The landmarks were measured again on both the reconstructed models and the physical RP models, and these were compared with those on dry skull. We focused on major sources of errors, such as CT scanning, conversion from CT data to STL models, and RP model fabrication. The results show that the overall error from skull to RP is 0.64±0.36mm(0.71±0.66%) in absolute value. This indicates that the RP technology can be acceptable in the real clinical applications. A clinical case that has applied RP models successfully for treatment planning and surgical rehearsal is presented. Although the use of RP models is rare in the medical area yet, we believe RP is promising in that it has a great potential in developing new tools which can aid diagnosis, treatment planning, surgical rehearsal, education, and so on. | en |
dc.language.iso | ko | en |
dc.publisher | 대한구강악안면외과학회 | en |
dc.subject | Rapid Prototyping | en |
dc.subject | 3D Medical Models | en |
dc.subject | Accuracy | en |
dc.title | Rapid Prototyping으로 제작한 3D Medical Model의 오차 측정에 관한 연구 (임상 적용 가능성 및 사례) | en |
dc.title.alternative | Accuracy tests of 3D rapid prototyping(RP) medical models: Its potential and clinical applications | en |
dc.type | Article | en |
dc.contributor.AlternativeAuthor | Choi, Jin-Young | - |
dc.contributor.AlternativeAuthor | Choi, Jung-Ho | - |
dc.contributor.AlternativeAuthor | Kim, Nam-Kuk | - |
dc.contributor.AlternativeAuthor | Lee, Jong-Ki | - |
dc.contributor.AlternativeAuthor | Kim, Myeng-Ki | - |
dc.contributor.AlternativeAuthor | Kim, Myung-Jin | - |
dc.contributor.AlternativeAuthor | Kim, Yeong-Ho | - |
- Appears in Collections:
- Files in This Item:
Item View & Download Count
Items in S-Space are protected by copyright, with all rights reserved, unless otherwise indicated.