Publications

Detailed Information

와우내 선형 전극의 위치와 삽입 손상에 대한 측두골 연구 : A Temporal Bone Study of Insertion Trauma and Intracochlear Position of Straight Type Electrode for Cochlear Implant

DC Field Value Language
dc.contributor.author박민현-
dc.contributor.author이호선-
dc.contributor.author박세익-
dc.contributor.author전상범-
dc.contributor.author김성준-
dc.contributor.author오승하-
dc.date.accessioned2009-08-27T03:18:59Z-
dc.date.available2009-08-27T03:18:59Z-
dc.date.issued2008-
dc.identifier.citationKorean J Otorhinolaryngol-Head Neck Surg 2008;51:999-1003en
dc.identifier.urihttps://hdl.handle.net/10371/7917-
dc.description.abstractBackground and Objectives:The goal of this study was to examine insertion depth, intracochlear position and insertion trauma
with the stimulation electrode of the Nurobiosys cochlear implant. Subjects and Method:Four electrodes were implanted
in fresh temporal bones of the human cadavers using realistic surgical procedures. Plain film X-ray images were taken from the
electrode inserted in the specimens to estimate the insertion depth. After the electrode implantation, all human temporal bones
were trimmed to extract the cochleae. The extracted cochleae from the temporal bone were immersed in acrylic resin to fix the
position of electrode placed in the scala tympani. The resin treated cochleae were cut in radial section and polished. All crosssections
were imaged with a microscope to assess the trauma by the electrode implantation. Results:The mean insertion depth
was about 300°with the cochlea angle. The insertion trauma was observed in one section of a temporal bone. The mean distance
from electrode to modiolus was about 0.75 millimeter. Conclusion:The incidence, severity of trauma and insertion depth of the
studied electrode showed similar results with that of other straight type electrode in literature.
en
dc.description.sponsorship전극을 제공한 Nurobiosys와 측두골 제작에 도움을 주신 Epstein
Laboratory at University of California측에도 감사를 드립니다.
en
dc.language.isokoen
dc.publisher대한이비인후과학회 = Korea Society of Otorhinolaryngology-Head and Neck Surgeryen
dc.subjectCochlear implanten
dc.subjectTemporal boneen
dc.subjectTraumaen
dc.subjectInsertion depthen
dc.title와우내 선형 전극의 위치와 삽입 손상에 대한 측두골 연구en
dc.title.alternativeA Temporal Bone Study of Insertion Trauma and Intracochlear Position of Straight Type Electrode for Cochlear Implanten
dc.typeArticleen
dc.contributor.AlternativeAuthorPark, Min-Hyun-
dc.contributor.AlternativeAuthorLee, Ho-Sun-
dc.contributor.AlternativeAuthorPark, Se-Ik-
dc.contributor.AlternativeAuthorJun, Sang Beom-
dc.contributor.AlternativeAuthorKim, Sung June-
dc.contributor.AlternativeAuthorOh, Seung-Ha-
dc.identifier.doi10.3342/kjorl-hns.2008.51.11.999-
Appears in Collections:
Files in This Item:

Altmetrics

Item View & Download Count

  • mendeley

Items in S-Space are protected by copyright, with all rights reserved, unless otherwise indicated.

Share