dc.description학위논문 (박사)-- 서울대학교 대학원 : 의과학과, 2015. 2. 이재성.-
dc.description.abstractIntroduction: A network based data acquisition (DAQ) system for positron emission tomography (PET) is developed. PET system with high sensitivity and resolution requires high performance DAQ system since large amount of data are being created. Most DAQ acquisition systems for PET are dedicated to one system in terms of output signal types, interface and channel configurations. Modular system which can adopt different input modules was introduced for easy configuration to receive different types of output signal from various PET systems. Field-programmable gate array (FPGA) based signal processing and data transfer with 1-Gbps Ethernet network was used for high system count rate and scalable system.
Methods: The DAQ system is composed of 3 boards, input, ADC and FPGA modules. Each DAQ system can acquire 40 analog signal inputs and 10 trigger inputs from PET system. 10-Gbps network switch is used for data transfer and scalable expansion of DAQ system.
Results: The 10-Gbps Ethernet network DAQ system was able to acquire the data from the PET scanner with maximum singles count rate of 57.6 Mcps.
Conclusions: A 10-Gbps Ethernet network based data acquisition system designed to acquire signals from various PET systems with high system count rate performance was developed. Scalablility from 40 to 10,000 readout channel is possible with use of Ethernet based DAQ system.
-
dc.description.tableofcontentsAbstract i
Contents ii
List of figures iv
List of tables vii
List of abbreviations viii
General Introduction 1
Chapter 1 3
High Performance Network based Data Acquisition System
Introduction 4
Material and Methods 7
Results 19
Discussion 24
Conclusion 26
Chapter 2 27
Analog Bipolar Multiplexing for DAQ Channel Expansion
Introduction 28
Material and Methods 31
Results 42
Discussion 45
Conclusion 49
Chapter 3 50
Application to the PET System for Simultaneous PET-MR Imaging
Introduction 51
Material and Methods 54
Results 63
Discussion 68
Conclusion 72
References 73
Abstract in Korean 79
-
dc.formatapplication/pdf-
dc.format.extent6465048 bytes-
dc.format.mediumapplication/pdf-
dc.language.isoen-
dc.publisher서울대학교 대학원-
dc.subject양전자방출단층촬영-
dc.subject데이터확득-
dc.subject.ddc610-
dc.titleNetwork based High Performance Data Acquisition System for PET Scanner-
dc.title.alternative양전자방출단층촬영을 위한 네트워크 기반 고성능 데이터획득 장치-
dc.typeThesis-
dc.contributor.AlternativeAuthorHyun Suk Yoon-
dc.description.degreeDoctor-
dc.citation.pagesix,80-
dc.contributor.affiliation의과대학 의과학과-
dc.date.awarded2015-02-
Appears in Collections:
Files in This Item:

Altmetrics

Item View & Download Count

  • mendeley

Items in S-Space are protected by copyright, with all rights reserved, unless otherwise indicated.

Share