dc.description.abstract | The purpose of this study was to evaluate the applicability of a rebound tonometer (TonoVet®) in pigeon eyes and to establish normal reference intraocular pressure (IOP) values in healthy pigeons. Twenty eyes of euthanized pigeons were used for the calibration of TonoVet® and 48 eyes of 24 adult pigeons for the reference IOP. First, IOP of pigeon eyes ex vivo were measured using the d and the p modes of the TonoVet® and compared with manometric IOP values from 5
to 80 mmHg. Then, to establish normal reference values, IOP was measured from clinically normal pigeons in vivo. Both the d and the p modes of the TonoVet® showed a strong linear correlation with the manometric IOP (R2 = 0.996 and 0.991, respectively). The obtained regression formulas were y = 0.431x1 + 2.154 and y = 0.330x2 - 0.673, respectively (x1, d mode of TonoVet® | - |
dc.description.abstract | x2, p mode of TonoVet® | - |
dc.description.abstract | y, manometric IOP). The d and the p modes steadily underestimated the actual IOP by approximately 1/2 and 1/3, respectively. Therefore, the formula obtained through the d mode was applied to obtain reference values. The calibrated IOP of normal pigeon eyes was 19.5 ± 4.4 mmHg. The actual IOP could be calculated using the presented formula. Considering the limitations of the p mode, it would be more appropriate to use the d mode. Therefore, the TonoVet® rebound tonometry under d mode is a reliable method for measuring IOP in pigeons. | - |
dc.description.tableofcontents | INTRODUCTION 1
MATERIALS AND METHODS 2 1. Experimental animals 2 2. Manomteric calibration of TonoVet® 3 3. Determination of normal pigeon IOP 6 4. Statistical analyses 8 RESULTS 9 1. Manomteric calibration of TonoVet® 9 2. Determination of normal pigeon IOP 13 DISCUSSIONS 15 CONCLUSIONS 18 REFERENCES 19 ABSTRACT IN KOREAN 23 | - |
dc.format | application/pdf | - |
dc.format.extent | 969102 bytes | - |
dc.format.medium | application/pdf | - |
dc.language.iso | en | - |
dc.publisher | 서울대학교 대학원 | - |
dc.subject | intraocular pressure | - |
dc.subject | rebound tonometry | - |
dc.subject | pigeons | - |
dc.subject | Columbiformes | - |
dc.subject | reference values | - |
dc.subject | linear regression analysis | - |
dc.subject.ddc | 636.089 | - |
dc.title | Intraocular Pressure Measurement with a Rebound Tonometer (TonoVet®) in Normal Pigeons | - |
dc.type | Thesis | - |
dc.description.degree | Master | - |
dc.contributor.affiliation | 수의과대학 수의학과 | - |
dc.date.awarded | 2017-08 | - |
- Appears in Collections:
- Files in This Item:
Item View & Download Count
Items in S-Space are protected by copyright, with all rights reserved, unless otherwise indicated.