Types and Arrangement of Thyroid Shields to Reduce Exposure of Surgeons to Ionizing Radiation During Intraoperative Use of C-arm Fluoroscopy

Cited 26 time in Web of Science Cited 32 time in Scopus
Authors

Lee, Seung Yeol; Min, Eungi; Bae, Jaekeon; Chung, Chin Youb; Lee, Kyoung Min; Kwon, Soon-Sun; Park, Moon Seok; Lee, Kisung

Issue Date
2013-11
Publisher
Lippincott Williams & Wilkins Ltd.
Citation
Spine, Vol.38 No.24, pp.2108-2112
Abstract
Study Design. Measurement of radiation dose from C-arm fluoroscopy during a simulated intraoperative use in spine surgery. Objective. To assess how the radiation dose is affected by changes in the types of thyroid shields used and by the arrangements or ways in which they are worn during the intraoperative use of C-arm fluoroscopy. Summary of Background Data. Although the danger to the thyroid from exposure to radiation is well known, there are no guidelines for the proper use of thyroid shields. Methods. Two photoluminescence dosimeters were used to measure the dose of scattered radiation arriving at the location of the thyroid in a whole-body phantom in the position of the surgeon. On an operating table beside this setup was an anthropomorphic chest phantom representing a patient for which treatment with C-arm fluoroscopy was simulated. Radiation doses were measured using 3 different arrangements of the thyroid shield: worn tightly, worn loosely, and worn loosely with a bismuth masking reagent. The same tests were performed using 2 kinds of thyroid shield: lead and lead-equivalent. Results. For the lead-shield group, radiation doses were measured in 3 arrangements; worn tightly, worn loosely, and worn loosely with a bismuth masking reagent, for which the results were 1.91 +/- 0.13, 2.35 +/- 0.22, and 1.86 +/- 0.13 mu Sv/min, respectively. Wearing the shield tight against the throat and wearing it loose with a bismuth masking reagent led to lower radiation exposure levels than by simply wearing the shield loosely (P <= 0.001). For the lead-equivalent shield group, doses were measured for the same 3 arrangements, for which the results were 1.79 +/- 0.12, 1.82 +/- 0.11, and 1.74 +/- 0.12 mu Sv/min. Lower scattered radiation doses were delivered to the thyroid in the lead-equivalent thyroid shield group compared with the lead thyroid shield group (P <= 0.001). The unshielded thyroid group received a radiation dose of 16.32 +/- 0.48 mu Sv/min. Conclusion. The use of some form of thyroid shield is essential during the use of C-arm fluoroscopy. It was found that the best way to reduce scattered radiation exposure to the thyroid was wearing the thyroid shield tightly or wearing it loosely in combination with a bismuth masking reagent.
ISSN
0362-2436
URI
https://hdl.handle.net/10371/192050
DOI
https://doi.org/10.1097/BRS.0b013e3182a8270d
Files in This Item:
There are no files associated with this item.
Appears in Collections:

Related Researcher

  • College of Medicine
  • Department of Medicine
Research Area Cerebral palsy, Motion analysis, Pediatric orthopedic surgery

Altmetrics

Item View & Download Count

  • mendeley

Items in S-Space are protected by copyright, with all rights reserved, unless otherwise indicated.

Share