악교정수술에서 광전자 포인트 마커를 이용한 상악골 위치 변화의 계측 및 계산 방법 연구 : Measurement and Algorithm Calculation of Maxillary Positioning Change by Use of an Optoelectronic Tracking System Marker in Orthognathic Surgery

Cited 0 time in Web of Science Cited 0 time in Scopus
Authors

박종웅; 김성민; 어미영; 박정민; 명훈; 이종호; 김명진

Issue Date
2011-05
Publisher
대한악안면성형재건외과학회
Citation
Maxillofacial Plastic Reconstructive Surgery, Vol.33 No.3, pp.233-240
Abstract
Purpose: To apply a computer assisted navigation system to orthognathic surgery, a simple and efficient measuring algorithm calculation based on affine transformation was designed. A method of improving accuracy and reducing errors in orthognathic surgery by use of an optical tracking camera was studied. Methods: A total of 5 points on one surgical splint were measured and tracked by the Polaris Vicra® (Northern Digital Inc Co., Ontario, Canada) optical tracking system in two cases. The first case was to apply the transformation matrix at pre- and postoperative situations, and the second case was to apply an affine transformation only after the postoperative situation. In each situation, the predictive measuring value was changed to the final measuring value via an affine transformation algorithm and the expected coordinates calculated from the model were compared with those of the patient in the operation room. Results: The mean measuring error was 1.027±0.587 using the affine transformation at pre- and postoperative situations and the average value after the postoperative situation was 0.928±0.549. The farther a coordinate region was from the reference coordinates which constitutes the transform matrixes, the bigger the measuring error was found which was calculated from an affine transformation algorithm. Conclusion: Most difference errors were brought from mainly measuring process and lack of reproducibility, the affine transformation algorithm formula from postoperative measuring values by using of optic tracking system between those of model surgery and those of patient surgery can be selected as minimizing the difference error. To reduce coordinate calculation errors, minimum transformation matrices must be used and reference points which determine an affine transformation must be close to the area where coordinates are measured and calculated, as well as the reference points need to be scattered.
ISSN
2288-8101
URI
https://hdl.handle.net/10371/207993
Files in This Item:
There are no files associated with this item.
Appears in Collections:

Related Researcher

  • School of Dentistry
  • Department of Dentistry
Research Area Dental implant related bony regeneration, Facial reconstruction, bioactive membrane & E-beam irradiation, 상악동 및 타액선 외과술, 임플란트 골재생술, 턱얼굴 미세수술재건

Altmetrics

Item View & Download Count

  • mendeley

Items in S-Space are protected by copyright, with all rights reserved, unless otherwise indicated.

Share