Publications

Detailed Information

A COMPARATIVE STUDY OF THE 1-PIECE AND 2-PIECE CONICAL ABUTMENT JOINT: THE STRENGTH AND THE FATIGUE RESISTANCE

DC Field Value Language
dc.contributor.authorKwon, Taek-Ka-
dc.contributor.authorYang, Jae-Ho-
dc.contributor.authorKim, Sung-Hun-
dc.contributor.authorHan, Jung-Suk-
dc.contributor.authorLee, Jai-Bong-
dc.date.accessioned2010-02-01T09:45:02Z-
dc.date.available2010-02-01T09:45:02Z-
dc.date.issued2007-
dc.identifier.citation2007;45(6)en
dc.identifier.issn0301-2875-
dc.identifier.urihttps://hdl.handle.net/10371/47670-
dc.description.abstractStatement of problem. The performance and maintenance of implant-supported prostheses are primarily dependent upon load transmission both at the bone-to-implant interface and within the implant-abutment-prosthesis complex. The design of the interface between components has been shown to have a profound influence on the stability of screw joints. Purpose. The Purpose of this study was to compare the strength and the fatigue resistance of 1-piece and 2-piece abutment connected to oral implant, utilizing an internal conical interface. Material and methods. Twenty tapered implants were embedded to the top of the fixture in acrylic resin blocks. Ten (1-piece) and (2-piece) abutments of the same dimension were assembled to the implant, respectively. The assembled units were mounted in a testing machine. A load was applied perpendicular to the long axis of the assemblies and the loading points was at the distance of 7mm from the block surface. Half of 1-piece and 2-piece abutment-implant units were tested for the evaluation of the bending strength, and the others were cyclically loaded for the evaluation of the fatigue resistance until plastic deformation occurred. Nonparametric statistical analysis was performed for the results. Results. Mean plastic and maximum bending moment were for the 1-piece abutment, and for the 2-piece abutment, respectively. Mean cycles and standard deviation when implant-abutment joint showed a first plastic deformation were . cycles for the 1-piece abutment and cycles for the 2-piece abutment. A 1-piece abutment showed significantly higher value than a 2-piece abutment in the first plastic bending moment (p<.05), maximum bending moment (p<.05) and fatigue strength (p<.05). Conclusion. Both 1-piece and 2-piece conical abutment had high strength and fatigue resistance and this suggests long-term durability without mechanical complication. However, the 1-piece conical abutment was more stable than the 2-piece conical abutment in the strength and the fatigue resistance.en
dc.language.isoenen
dc.publisher대한치과보철학회en
dc.subjectConical jointen
dc.subjectInterfaceen
dc.subjectBending strengthen
dc.subjectFatigue resistanceen
dc.titleA COMPARATIVE STUDY OF THE 1-PIECE AND 2-PIECE CONICAL ABUTMENT JOINT: THE STRENGTH AND THE FATIGUE RESISTANCEen
dc.typeArticleen
dc.contributor.AlternativeAuthor권택가-
dc.contributor.AlternativeAuthor양재호-
dc.contributor.AlternativeAuthor김성훈-
dc.contributor.AlternativeAuthor한중석-
dc.contributor.AlternativeAuthor이재봉-
Appears in Collections:
Files in This Item:

Altmetrics

Item View & Download Count

  • mendeley

Items in S-Space are protected by copyright, with all rights reserved, unless otherwise indicated.

Share