A Postoperative Ventilatory Failure : 수술후 발생한 환기 부전증

DC Field Value Language
dc.contributor.authorKim, Seang Deok-
dc.contributor.authorSung, Sea Wook-
dc.contributor.authorSuh, Kyung Phill-
dc.date.accessioned2009-08-07-
dc.date.available2009-08-07-
dc.date.issued1985-09-
dc.identifier.citationSeoul J Med, Vol.26 No.3, pp. 283-286-
dc.identifier.issn0582-6802-
dc.identifier.urihttps://hdl.handle.net/10371/6404-
dc.description.abstractA case of severe ventilatory failure with PaCOz, as high as 197 torr is reported
postoperatively in a 3-year-old male patient who has received open heart surgery for total
correction of known TOF (Tetralogy of Fallot). Many procedures were tried to treat the
ventilatory failure but failed. Eventually we noticed malfunctioning of the humidifier assembled
in Bear Cub infant pressure-limited ventilator. Vigorous endotracheal suction combined with
saline irrigation improved the ventilatory failure successfully.
-
dc.language.isoen-
dc.publisherSeoul National University College of Medicine-
dc.subjectVentilatory insufficiency-
dc.subjectTOF-
dc.subjectHumidifier-
dc.subjectRespiratory care-
dc.titleA Postoperative Ventilatory Failure-
dc.title.alternative수술후 발생한 환기 부전증-
dc.typeSNU Journal-
dc.contributor.AlternativeAuthor김성덕-
dc.contributor.AlternativeAuthor성시욱-
dc.contributor.AlternativeAuthor서경필-
dc.citation.journaltitle서울 의대 잡지-
dc.citation.journaltitle서울 의대 학술지-
dc.citation.journaltitleSeoul Journal of Medicine-
dc.citation.endpage286-
dc.citation.number3-
dc.citation.pages283-286-
dc.citation.startpage283-
dc.citation.volume26-
Appears in Collections:
Files in This Item:

Altmetrics

Item View & Download Count

  • mendeley

Items in S-Space are protected by copyright, with all rights reserved, unless otherwise indicated.

Share