Publications

Detailed Information

Development of Polymer Prosthetic Heart Valve - Fabrication and In Vitro Test- : 고분자 인공 심장판막의 개발 -제작과 성능실험

DC Field Value Language
dc.contributor.authorKim, In Young-
dc.contributor.authorKim, Jong Won-
dc.contributor.authorJung, Dae Young-
dc.contributor.authorKim, Cheol Sang-
dc.contributor.authorMin, Byoung Goo-
dc.date.accessioned2009-08-09T23:11:51Z-
dc.date.available2009-08-09T23:11:51Z-
dc.date.issued1991-03-
dc.identifier.citationSeoul J Med, Vol.32 No.1, pp. 35-42-
dc.identifier.issn0582-6802-
dc.identifier.urihttps://hdl.handle.net/10371/6709-
dc.description.abstractProsthetic heart valves (PHV) with various types and materials have been
developed and used until now, but PHVs have various problems that need improving.
A new type of polymer PHV have been developed to ameliorate these problems. A
hydrodynamic tester has also been developed for the precise evaluation of hydrodynamic
performance of the developed PHV.
PHV are composed of polyurethane, a material commonly used in biomedical materials
because of its good biocompatibility and mechanical properties. To investigate
its hydrodynamic properties, a trileaflet and bileaflet type polymer valve was fabricated.
The results of comparative tests with the 2 polymer valves and a mechanical valve
showed that the trileaflet polymer valve had a better hydrodynamic performance than
the others. Though the bileaflet polymer valve showed a higher pressure drop and regurgitation
than the trileaflet valve due to thicker leaflet flection of the bileaflet valve,
if its fabrication method and design were corrected, a better result could be expected.
For a study of the long-term stability of the PHV, accelerated endurance tests and
stress analysis of PHVs are needed.
-
dc.language.isoen-
dc.publisherSeoul National University College of Medicine-
dc.subjectProsthetic heart valve-
dc.subjectHydrodynamic tester-
dc.subjectTrileaflet polymer valve-
dc.subjectBileaflet polymer valve-
dc.titleDevelopment of Polymer Prosthetic Heart Valve - Fabrication and In Vitro Test--
dc.title.alternative고분자 인공 심장판막의 개발 -제작과 성능실험-
dc.typeSNU Journal-
dc.contributor.AlternativeAuthor김인영-
dc.contributor.AlternativeAuthor김종원-
dc.contributor.AlternativeAuthor정대영-
dc.contributor.AlternativeAuthor김철생-
dc.contributor.AlternativeAuthor민병구-
dc.citation.journaltitle서울 의대 잡지-
dc.citation.journaltitle서울 의대 학술지-
dc.citation.journaltitleSeoul Journal of Medicine-
dc.citation.endpage42-
dc.citation.number1-
dc.citation.pages35-42-
dc.citation.startpage35-
dc.citation.volume32-
Appears in Collections:
Files in This Item:

Altmetrics

Item View & Download Count

  • mendeley

Items in S-Space are protected by copyright, with all rights reserved, unless otherwise indicated.

Share