Publications

Detailed Information

A Four-Layer DOI Detector With a Relative Offset for Use in an Animal PET System

DC Field Value Language
dc.contributor.authorIto, Mikiko-
dc.contributor.authorLee, Jae Sung-
dc.contributor.authorKwon, Sun Il-
dc.contributor.authorLee, Geon Song-
dc.contributor.authorLee, Kyong Sei-
dc.contributor.authorLee, Seok Jae-
dc.contributor.authorHong, Seong Jong-
dc.contributor.authorRhee, June Tak-
dc.contributor.authorSim, Kwang-Souk-
dc.contributor.authorHong, Byungsik-
dc.date.accessioned2012-06-29T02:15:27Z-
dc.date.available2012-06-29T02:15:27Z-
dc.date.issued2010-06-
dc.identifier.citationIEEE TRANSACTIONS ON NUCLEAR SCIENCE; Vol.57 3; 976-981ko_KR
dc.identifier.issn0018-9499-
dc.identifier.urihttps://hdl.handle.net/10371/77886-
dc.description.abstractFor animal PET systems to achieve high sensitivity without adversely affecting spatial resolution, they must have the ability to measure depth-of-interaction (DOI). In this paper, we propose a novel four-layer PET system, and present the performances of modules built to verify the concept of the system. Each layer in the four-layer PET system has a relative offset of half a crystal pitch from other layers. Performances of the four-layer detector were estimated using a GATE Monte Carlo simulation code. The proposed system consists of six H9500 PMTs, each of which contains 3193 crystals. A sensitivity of 11.8% was obtained at the FOV center position of the proposed system. To verify the concept, we tested a PET module constructed using a H9500 flat panel PMT and LYSO crystals of cross-sectional area 1.5 x 1.5 mm(2). The PET module was irradiated with a 1.8 MBq (22)Na radiation source from the front or side of the crystals to obtain flood images of each crystal. Collimation for side irradiation was achieved using a pair of lead blocks of dimension 50 100 200 mm(3). All crystals in the four layers were clearly identified in flood images, thus verifying the DOI capability of the proposed four-layer PET system. We also investigated the optimal combination of crystal lengths in the four-layer PET system using the GATE Monte Carlo simulation code to generate events from simulated radiation sources, and using the ML-EM algorithm to reconstruct simulated radiation sources. The combination of short crystal lengths near radiation sources and long crystal lengths near the PMT provides better spatial resolution than combinations of same crystal lengths in the four-layer PET system.ko_KR
dc.language.isoenko_KR
dc.publisherIEEE-INST ELECTRICAL ELECTRONICS ENGINEERS INCko_KR
dc.subjectDepth of interaction (DOI))ko_KR
dc.subjectH9500 photomultiplier tube (PMT)ko_KR
dc.subjectGATE Monte Carlo simulationko_KR
dc.subjectfour-layer animal PETko_KR
dc.titleA Four-Layer DOI Detector With a Relative Offset for Use in an Animal PET Systemko_KR
dc.typeArticleko_KR
dc.contributor.AlternativeAuthor이재성-
dc.contributor.AlternativeAuthor권선일-
dc.contributor.AlternativeAuthor이건송-
dc.contributor.AlternativeAuthor홍병식-
dc.contributor.AlternativeAuthor이경세-
dc.contributor.AlternativeAuthor심광숙-
dc.contributor.AlternativeAuthor이석재-
dc.contributor.AlternativeAuthor이준탁-
dc.contributor.AlternativeAuthor홍승종-
dc.identifier.doi10.1109/TNS.2010.2044892-
dc.citation.journaltitleIEEE TRANSACTIONS ON NUCLEAR SCIENCE-
dc.description.citedreferenceHong SJ, 2008, IEEE T NUCL SCI, V55, P912, DOI 10.1109/TNS.2008.920258-
dc.description.citedreferenceKWON SI, 2008, NUCL MED MOL IMAGING, V42, P469-
dc.description.citedreference*NEMA STAND, 2008, PUBL NU-
dc.description.citedreferenceKim JS, 2007, J NUCL MED, V48, P1527, DOI 10.2967/jnumed.107.040550-
dc.description.citedreferenceInadama N, 2006, IEEE T NUCL SCI, V53, P30, DOI 10.1109/TNS.2005.862963-
dc.description.citedreferencePomper MG, 2005, CURR PHARM DESIGN, V11, P3247-
dc.description.citedreferenceJan S, 2004, PHYS MED BIOL, V49, P4543, DOI 10.1088/0031-9155/49/19/007-
dc.description.citedreferenceJagoda EM, 2004, NUCL MED BIOL, V31, P771, DOI 10.1016/j.nucmedbio.2004.04.003-
dc.description.citedreferenceInadama N, 2004, IEEE T NUCL SCI, V51, P58, DOI 10.1109/TNS.2004.823332-
dc.description.citedreferenceZhang N, 2003, IEEE T NUCL SCI, V50, P1398, DOI 10.1109/TNS.2003.817954-
dc.description.citedreferenceYamaya T, 2003, IEEE T NUCL SCI, V50, P1404, DOI 10.1109/TNS.2003.817307-
dc.description.citedreferenceDaube-Witherspoon ME, 2002, J NUCL MED, V43, P1398-
dc.description.citedreferenceSTREUN M, 2002, IEEE NSS MIC C REC, V3, P1636-
dc.description.citedreferenceShao Y, 2000, IEEE T NUCL SCI, V47, P1051, DOI 10.1109/23.856546-
dc.description.citedreferenceSeidel J, 1999, IEEE T NUCL SCI, V46, P485, DOI 10.1109/23.775567-
dc.description.citedreferenceMacDonald LR, 1998, IEEE T NUCL SCI, V45, P2232, DOI 10.1109/23.708354-
dc.description.citedreferenceMurayama H, 1998, IEEE T NUCL SCI, V45, P1152, DOI 10.1109/23.681994-
dc.description.citedreferenceYamamoto S, 1998, IEEE T NUCL SCI, V45, P1078, DOI 10.1109/23.681982-
dc.description.citedreferenceDahlbom M, 1997, IEEE T NUCL SCI, V44, P1114, DOI 10.1109/23.596974-
dc.description.citedreferenceMOSES WW, 1995, IEEE T NUCL SCI, V42, P1085, DOI 10.1109/23.467744-
dc.description.citedreferenceHOFFMAN EJ, 1989, IEEE T NUCL SCI, V36, P1108, DOI 10.1109/23.34613-
dc.description.citedreferenceDAUBEWITHERSPOON ME, 1987, J NUCL MED, V28, P1717-
dc.description.tc4-