Publications
Detailed Information
A Four-Layer DOI Detector With a Relative Offset for Use in an Animal PET System
DC Field | Value | Language |
---|---|---|
dc.contributor.author | Ito, Mikiko | - |
dc.contributor.author | Lee, Jae Sung | - |
dc.contributor.author | Kwon, Sun Il | - |
dc.contributor.author | Lee, Geon Song | - |
dc.contributor.author | Lee, Kyong Sei | - |
dc.contributor.author | Lee, Seok Jae | - |
dc.contributor.author | Hong, Seong Jong | - |
dc.contributor.author | Rhee, June Tak | - |
dc.contributor.author | Sim, Kwang-Souk | - |
dc.contributor.author | Hong, Byungsik | - |
dc.date.accessioned | 2012-06-29T02:15:27Z | - |
dc.date.available | 2012-06-29T02:15:27Z | - |
dc.date.issued | 2010-06 | - |
dc.identifier.citation | IEEE TRANSACTIONS ON NUCLEAR SCIENCE; Vol.57 3; 976-981 | ko_KR |
dc.identifier.issn | 0018-9499 | - |
dc.identifier.uri | https://hdl.handle.net/10371/77886 | - |
dc.description.abstract | For animal PET systems to achieve high sensitivity without adversely affecting spatial resolution, they must have the ability to measure depth-of-interaction (DOI). In this paper, we propose a novel four-layer PET system, and present the performances of modules built to verify the concept of the system. Each layer in the four-layer PET system has a relative offset of half a crystal pitch from other layers. Performances of the four-layer detector were estimated using a GATE Monte Carlo simulation code. The proposed system consists of six H9500 PMTs, each of which contains 3193 crystals. A sensitivity of 11.8% was obtained at the FOV center position of the proposed system. To verify the concept, we tested a PET module constructed using a H9500 flat panel PMT and LYSO crystals of cross-sectional area 1.5 x 1.5 mm(2). The PET module was irradiated with a 1.8 MBq (22)Na radiation source from the front or side of the crystals to obtain flood images of each crystal. Collimation for side irradiation was achieved using a pair of lead blocks of dimension 50 100 200 mm(3). All crystals in the four layers were clearly identified in flood images, thus verifying the DOI capability of the proposed four-layer PET system. We also investigated the optimal combination of crystal lengths in the four-layer PET system using the GATE Monte Carlo simulation code to generate events from simulated radiation sources, and using the ML-EM algorithm to reconstruct simulated radiation sources. The combination of short crystal lengths near radiation sources and long crystal lengths near the PMT provides better spatial resolution than combinations of same crystal lengths in the four-layer PET system. | ko_KR |
dc.language.iso | en | ko_KR |
dc.publisher | IEEE-INST ELECTRICAL ELECTRONICS ENGINEERS INC | ko_KR |
dc.subject | Depth of interaction (DOI)) | ko_KR |
dc.subject | H9500 photomultiplier tube (PMT) | ko_KR |
dc.subject | GATE Monte Carlo simulation | ko_KR |
dc.subject | four-layer animal PET | ko_KR |
dc.title | A Four-Layer DOI Detector With a Relative Offset for Use in an Animal PET System | ko_KR |
dc.type | Article | ko_KR |
dc.contributor.AlternativeAuthor | 이재성 | - |
dc.contributor.AlternativeAuthor | 권선일 | - |
dc.contributor.AlternativeAuthor | 이건송 | - |
dc.contributor.AlternativeAuthor | 홍병식 | - |
dc.contributor.AlternativeAuthor | 이경세 | - |
dc.contributor.AlternativeAuthor | 심광숙 | - |
dc.contributor.AlternativeAuthor | 이석재 | - |
dc.contributor.AlternativeAuthor | 이준탁 | - |
dc.contributor.AlternativeAuthor | 홍승종 | - |
dc.identifier.doi | 10.1109/TNS.2010.2044892 | - |
dc.citation.journaltitle | IEEE TRANSACTIONS ON NUCLEAR SCIENCE | - |
dc.description.citedreference | Hong SJ, 2008, IEEE T NUCL SCI, V55, P912, DOI 10.1109/TNS.2008.920258 | - |
dc.description.citedreference | KWON SI, 2008, NUCL MED MOL IMAGING, V42, P469 | - |
dc.description.citedreference | *NEMA STAND, 2008, PUBL NU | - |
dc.description.citedreference | Kim JS, 2007, J NUCL MED, V48, P1527, DOI 10.2967/jnumed.107.040550 | - |
dc.description.citedreference | Inadama N, 2006, IEEE T NUCL SCI, V53, P30, DOI 10.1109/TNS.2005.862963 | - |
dc.description.citedreference | Pomper MG, 2005, CURR PHARM DESIGN, V11, P3247 | - |
dc.description.citedreference | Jan S, 2004, PHYS MED BIOL, V49, P4543, DOI 10.1088/0031-9155/49/19/007 | - |
dc.description.citedreference | Jagoda EM, 2004, NUCL MED BIOL, V31, P771, DOI 10.1016/j.nucmedbio.2004.04.003 | - |
dc.description.citedreference | Inadama N, 2004, IEEE T NUCL SCI, V51, P58, DOI 10.1109/TNS.2004.823332 | - |
dc.description.citedreference | Zhang N, 2003, IEEE T NUCL SCI, V50, P1398, DOI 10.1109/TNS.2003.817954 | - |
dc.description.citedreference | Yamaya T, 2003, IEEE T NUCL SCI, V50, P1404, DOI 10.1109/TNS.2003.817307 | - |
dc.description.citedreference | Daube-Witherspoon ME, 2002, J NUCL MED, V43, P1398 | - |
dc.description.citedreference | STREUN M, 2002, IEEE NSS MIC C REC, V3, P1636 | - |
dc.description.citedreference | Shao Y, 2000, IEEE T NUCL SCI, V47, P1051, DOI 10.1109/23.856546 | - |
dc.description.citedreference | Seidel J, 1999, IEEE T NUCL SCI, V46, P485, DOI 10.1109/23.775567 | - |
dc.description.citedreference | MacDonald LR, 1998, IEEE T NUCL SCI, V45, P2232, DOI 10.1109/23.708354 | - |
dc.description.citedreference | Murayama H, 1998, IEEE T NUCL SCI, V45, P1152, DOI 10.1109/23.681994 | - |
dc.description.citedreference | Yamamoto S, 1998, IEEE T NUCL SCI, V45, P1078, DOI 10.1109/23.681982 | - |
dc.description.citedreference | Dahlbom M, 1997, IEEE T NUCL SCI, V44, P1114, DOI 10.1109/23.596974 | - |
dc.description.citedreference | MOSES WW, 1995, IEEE T NUCL SCI, V42, P1085, DOI 10.1109/23.467744 | - |
dc.description.citedreference | HOFFMAN EJ, 1989, IEEE T NUCL SCI, V36, P1108, DOI 10.1109/23.34613 | - |
dc.description.citedreference | DAUBEWITHERSPOON ME, 1987, J NUCL MED, V28, P1717 | - |
dc.description.tc | 4 | - |
- Appears in Collections:
- Files in This Item:
- There are no files associated with this item.
Item View & Download Count