Publications

Detailed Information

Efficacy and safety of a new elastic tourniquet cuff in total knee arthroplasty: a prospective randomized controlled study

Cited 4 time in Web of Science Cited 7 time in Scopus
Authors

Lee, O-Sung; Lee, Myung Chul; Han, Hyuk-Soo

Issue Date
2017-08-08
Publisher
BioMed Central
Citation
BioMedical Engineering OnLine, 16(1):102
Keywords
Total knee arthroplastyTourniquetAutomatic pneumatic tourniquetCuff
Description
Abbreviations
LOP: limb occlusion pressure; RTP: recommended tourniquet pressure; DVT: deep vein thrombosis; TKA: total knee arthroplasty; ROM: range
Abstract
Abstract

Background
The effects of cuff shape, timing of tourniquet application, and automated systems using limb occlusion pressure (LOP) have been reported to minimize the appropriate tourniquet pressure. However, studies on the raw material of the cuff itself to reduce the complicationsrelated to the tourniquet have been very rarely reported. The purpose of this study is to report the efficacy and safety of a tourniquet system with a new elastic cuff in which pressure is set with LOP in total knee arthroplasty (TKA).

Methods
A total of 63 patients who underwent primary TKA for osteoarthritis were enrolled from July to December 2015. Thirty-one patients were allocated to the new elastic cuff group and 32 in the conventional cuff group. Bloodless surgical field, pain visual analog scale (VAS) on the thigh, thigh circumference, range of motion, incidence of deep vein thrombosis, and muscle enzyme level after surgery were checked and compared between the 2 groups.

Results
Only 1 of the 31 patients in the elastic cuff group required more pressure for obtaining a bloodless surgical field, whereas 4 of the 32 patients in the conventional cuff group required more pressure to complete surgery without being disturbed by sustained bleeding. Two patients in the conventional cuff group needed treatment for blisters and bullae at the tourniquet application site. There was no difference in pain VAS score, thigh circumference, range of motion, incidence of deep vein thrombosis, and level of muscle enzyme.

Conclusions
A new elastic tourniquet cuff provided a more proper bloodless surgical field with less adjustment of tourniquet pressure despite a similar level of tourniquet pressure compared to the conventional cuff and had a low incidence of skin complications on the site of tourniquet application in TKA. These benefits make it an effective and safe medical device for orthopedic surgery requiring a tourniquet, such as TKA.
ISSN
1475-925X
Language
English
URI
https://hdl.handle.net/10371/138308
DOI
https://doi.org/10.1186/s12938-017-0393-0
Files in This Item:
Appears in Collections:

Altmetrics

Item View & Download Count

  • mendeley

Items in S-Space are protected by copyright, with all rights reserved, unless otherwise indicated.

Share