Publications

Detailed Information

Design modifications to Rotating Modulation Collimators for extended field-of-view

Cited 2 time in Web of Science Cited 3 time in Scopus
Authors

Barros, S.; Kim, G.

Issue Date
2021-04
Publisher
Pergamon Press Ltd.
Citation
Applied Radiation and Isotopes, Vol.170, p. 109597
Abstract
Rotating Modulation Collimators (RMC) are devices that can be used to determine the location of a radiation source. In this work, a design with extended field-of-view (FOV) is proposed to allow faster measurements by providing a wider scanned area. Monte Carlo simulations were used to calculate the response of the RMC system, system matrix and modulation profiles. The Maximum Likelihood Expectation Maximization algorithm was used for image reconstruction. The proposed design provides a FOV up to 60 degrees with a maximum error of 5 degrees or up to 70 degrees with a maximum error of 15 degrees.
ISSN
0969-8043
URI
https://hdl.handle.net/10371/180013
DOI
https://doi.org/10.1016/j.apradiso.2021.109597
Files in This Item:
There are no files associated with this item.
Appears in Collections:

Altmetrics

Item View & Download Count

  • mendeley

Items in S-Space are protected by copyright, with all rights reserved, unless otherwise indicated.

Share