Publications

Detailed Information

Scanning methodology for contact lens-type ocular in vivo dosimeter (CLOD) dosimetry applying a silicone material

Cited 0 time in Web of Science Cited 0 time in Scopus
Authors

Son, Jaeman; Cho, Jin Dong; Jung, Seongmoon; Choi, Chang Heon; Park, Jong Min; Kim, Jung-in

Issue Date
2022-05
Publisher
BioMed Central
Citation
Radiation Oncology, Vol.17 No.1, p. 88
Abstract
Purpose Contact lens-type ocular in vivo dosimeters (CLODs) were recently developed as the first in vivo dosimeter that can be worn directly on the eye to measure the dose delivered to the lens during radiotherapy. However, it has an inherent uncertainty because of its curved shape. Newton's ring effect inevitably occurs because the spacing between the glass window and the active layer is not constant. Furthermore, it involves a large uncertainty because the objective of the CLOD with such morphological characteristics is to measure the dose delivered to an out-of-field lens. In this study, we aimed to investigate the effects of various compensating materials on the sensitivity, accuracy, and uniformity of analysis using a curved CLOD. We developed a new scanning methodology that involves applying a compensating material to reduce the uncertainty caused by the air gap. Methods Four compensating materials-Dragon Skin (TM) 10 (DS), a transparent silicon material, SORTA-Clear (TM) 40 (SC), optical grease (OG), and air (no compensating material)-were used in this study. The CLOD was scanned in the reflective mode and transmission mode using each compensating material. We then examined the sensitivity, accuracy, and scan uniformity to evaluate the scanning methodology using compensating materials. Results The increase in sensitivity was the highest for OG compared to that for air in the reflective mode. On average, the sensitivity in the reflective mode was higher than that in the transmission mode by a factor of 2.5 for each dose. Among the four compensating materials, OG had the smallest uncertainty. Therefore, the best scan uniformity was achieved when OG was used. Conclusions Scanning methodology was proposed in which a compensating material is applied for a curved lens-type dosimeter. Our results show that OG is the most suitable compensating material to obtain the best accuracy of dose analysis. Following this methodology, the scan uncertainty of curved dosimeters significantly decreased.
ISSN
1748-717X
URI
https://hdl.handle.net/10371/182594
DOI
https://doi.org/10.1186/s13014-022-02056-9
Files in This Item:
There are no files associated with this item.
Appears in Collections:

Altmetrics

Item View & Download Count

  • mendeley

Items in S-Space are protected by copyright, with all rights reserved, unless otherwise indicated.

Share