Measurements of surgeons' exposure to ionizing radiation dose: comparison of conventional and mini C-arm fluoroscopy

Cited 11 time in Web of Science Cited 15 time in Scopus
Authors

Sung, K. H.; Min, E.; Chung, C. Y.; Jo, B. C.; Park, M. S.; Lee, K.

Issue Date
2016-03
Publisher
SAGE Publications
Citation
Journal of Hand Surgery: European Volume, Vol.41 No.3, pp.340-345
Abstract
This study was performed to measure the equivalent scattered radiation dose delivered to susceptible organs while simulating orthopaedic surgery using conventional and mini C-arm fluoroscopy. In addition, shielding effects on the thyroid, thymus, and gonad, and the direct exposure delivered to the patient's hands were also compared. A conventional and mini C-arms were installed in an operating room, and a hand and an operator phantom were used to simulate a patient's hand and a surgeon. Photoluminescence dosimeters were used to measure the equivalent dose by scattered radiation arriving at the thyroid, thymus, and gonad on a whole-body phantom in the position of the surgeon. Equivalent scattered radiation doses were measured in four groups: (1) unshielded conventional C-arm group; (2) unshielded mini C-arm group; (3) lead-shielded conventional C-arm group; and (4) lead-shielded mini C-arm group. Equivalent scattered radiation doses to the unshielded group were significantly lower in the mini C-arm group than those in the conventional C-arm group for all organs. The gonad in the lead-shielded conventional C-arm group showed the highest equivalent dose among operator-susceptible organs, and radiation dose was reduced by approximately 96% compared with that in the unshielded group. Scattered radiation was not detected in any susceptible organ in the lead-shielded mini C-arm group. The direct radiation dose to the hand phantom measured from the mini C-arm was significantly lower than that measured from the conventional C-arm. The results show that the equivalent scattered radiation dose to the surgeon's susceptible organs and the direct radiation dose to a patient's hand can be decreased significantly by using a mini C-arm rather than a conventional C-arm. However, protective lead garments, such as a thyroid shield and apron, should be applied to minimize radiation exposure to susceptible organs, even during use of mini C-arm fluoroscopy.
ISSN
1753-1934
URI
https://hdl.handle.net/10371/191982
DOI
https://doi.org/10.1177/1753193415590388
Files in This Item:
There are no files associated with this item.
Appears in Collections:

Related Researcher

  • College of Medicine
  • Department of Medicine
Research Area Cerebral palsy, Motion analysis, Pediatric orthopedic surgery