임플랜트 지대주의 삭제과정이 결합부 안정성에 미치는 영향 : THE EFFECT OF PREPARATION PROCEDURE ON IMPLANT-ABUTMENT JOINT STABILITY
DC Field | Value | Language |
---|---|---|
dc.contributor.author | 이장욱 | - |
dc.contributor.author | 김창회 | - |
dc.contributor.author | 장경수 | - |
dc.contributor.author | 임영준 | - |
dc.date.accessioned | 2010-02-01T09:40:18Z | - |
dc.date.available | 2010-02-01T09:40:18Z | - |
dc.date.issued | 2005 | - |
dc.identifier.citation | 2005;43(5) | en |
dc.identifier.issn | 0301-2875 | - |
dc.identifier.uri | https://hdl.handle.net/10371/47574 | - |
dc.description.abstract | Statement of problem: Little is known about the effect of abutment preparation procedure on do-torque values in different implant platform and the relationship of final do-torque values with different implant platform size. Purpose: This study evaluated the effect of abutment preparation procedure on do-torque values in different implant platform and the relationship of final do-torque values with different implant platform size. Material and method: Six ITI implants (2 narrow-neck implants, 2 regular-neck implants, 2 wide-neck implants) and six Branemark implants (2 narrow platforms, 2 regular platforms, 2 wide platforms) were embedded in each acrylic resin block with epoxy resin. Eighteen abutments (6 narrow-neck implant-abutments, 6 regular-neck implant-abutments, 6 wide-neck implant-abutments) and eighteen esthetic abutments (6 narrow platform-abutments, 6 regular platform-abutments, 6 wide platform-abutments) were tightened to each implant with digital torque gauge. Initial do-torque values were measured using digital torque gauge. After preparation of abutments, Final do-torque values were measured with digital torque gauge. Results and conclusion: 1. Screws loosening or abutments motion were not detected in all experimental group, but some scratches of implant-abutment joints were detected in all group 2. Reduction ratios of final do-torque values were greater than initial do-torque values in all measured group, except in narrow-neck implant-abutment group (p<0.05). 3. Reduction ratios of final do-torque values in wide-neck implant-abutment group were greater than regular-neck implant-abutment group (p<0.01). 4. The greatest standard deviation value was detected in wide platform group in both implant systems. | en |
dc.language.iso | ko | en |
dc.publisher | 대한치과보철학회 | en |
dc.subject | Preparable abutment | en |
dc.subject | Initial do-torque value | en |
dc.subject | Final do-torque value | en |
dc.subject | Screw loosening | en |
dc.subject | Joint stability | en |
dc.title | 임플랜트 지대주의 삭제과정이 결합부 안정성에 미치는 영향 | en |
dc.title.alternative | THE EFFECT OF PREPARATION PROCEDURE ON IMPLANT-ABUTMENT JOINT STABILITY | en |
dc.type | Article | en |
dc.contributor.AlternativeAuthor | Lee, Jang-Wook | - |
dc.contributor.AlternativeAuthor | Kim, Chang-Whe | - |
dc.contributor.AlternativeAuthor | Jang, Kyung-Soo | - |
dc.contributor.AlternativeAuthor | Lim, Young-Jun | - |
- Appears in Collections:
Item View & Download Count
Items in S-Space are protected by copyright, with all rights reserved, unless otherwise indicated.