smartpegTM의 고압멸균소독 및 재사용이 임플랜트 안정성 지수(ISQ) 측정에 미치는 영향 : THE EFFECT OF AUTOCLAVE STERILIZATION AND REUSE OF SMARTPEGtm ON THE IMPLANT STABILITY QUOTIENT (ISQ) MEASUREMENT

DC Field Value Language
dc.contributor.authorKang, In-Ho-
dc.contributor.authorKim, Myung-Joo-
dc.contributor.authorLim, Young-Jun-
dc.contributor.authorKim, Chang-Whe-
dc.date.accessioned2010-02-01T09:58:04Z-
dc.date.available2010-02-01T09:58:04Z-
dc.date.issued2007-
dc.identifier.citation대한치과보철학회지 2007;45(5)en
dc.identifier.issn0301-2875-
dc.identifier.urihttps://hdl.handle.net/10371/47932-
dc.description.abstractStatement of problem: Resonance frequency analysis is evaluated as the non-invasive and objective method for the evaluation of implant stability and has been increasingly used. It is necessary to evaluate the factors affect the ISQ measurement stability. Purpose: This study was performed to evaluate the effect of the autoclave sterilization and reuse of on ISQ measurement. Material and methods: (Integration Diagnostics Ltd., , Sweden) of autoclave group (A) was autoclave sterilized 9 times and of reuse group (R) was reused 9 times. Ten were allocated to each group and after each autoclave sterilization and reuse, implant stability quotient (ISQ) values were measured 3 times from the two directions a and b at a right angle. mentor (Integration Diagnostics Ltd. , Sweden) was used and type 1 (article no. 100353) was selected according to reference list. Osstem Implant US II future (Osstem Co., Seoul, Korea) in was embedded in the self-curing acrylic resin (, Lang Dental, U.S.A.). Data was statistically analyzed by one-way ANOVA and scheffe test was done where a significant difference exist. Correlation test was also done between ISQ value and the number of autoclave sterilization or reuse. Results: 1. In autoclave group, the means and sd. of ISQ value before autoclave sterilization were at direction a and b. There was significant differences between autoclave groups at direction a and b (P=.000). 2. In reuse group, the means and sd. of ISQ value before reuse were at direction a and b. There was no significant difference between reuse groups at direction a and b (P>.05). 3. There was a weak positive correlation between the number of reuse and ISQ value at direction a and b ( and .246, P<.01). Conclusion: Within the limitations of this study, the following conclusions were drawn. Till ninth reuse of , the ISQ measurement stability did not be affected. After twice autoclave sterilization of the ISQ measurement stability was affected.en
dc.description.sponsorship이 연구는 서울대학교 신임교수 연구정착금으로 지원되는 연구비에 의하여 수행되었음.en
dc.language.isokoen
dc.publisher대한치과보철학회en
dc.subjectsmartpeg TMen
dc.subjectAutoclave Sterilizationen
dc.subjectReuseen
dc.subjectISQ valueen
dc.subjectResonance Frequency Analen
dc.titlesmartpegTM의 고압멸균소독 및 재사용이 임플랜트 안정성 지수(ISQ) 측정에 미치는 영향en
dc.title.alternativeTHE EFFECT OF AUTOCLAVE STERILIZATION AND REUSE OF SMARTPEGtm ON THE IMPLANT STABILITY QUOTIENT (ISQ) MEASUREMENTen
dc.typeArticleen
dc.contributor.AlternativeAuthor강인호-
dc.contributor.AlternativeAuthor김창회-
dc.contributor.AlternativeAuthor임영준-
dc.contributor.AlternativeAuthor김명주-
Appears in Collections:
Files in This Item:

Altmetrics

Item View & Download Count

  • mendeley

Items in S-Space are protected by copyright, with all rights reserved, unless otherwise indicated.

Share