Comparison of the color of natural teeth measured by a colorimeter and Shade Vision System

DC Field Value Language
dc.contributor.authorCho, Byeong-Hoon-
dc.contributor.authorLim, Yong-Kyu-
dc.contributor.authorLee, Yong-Keun-
dc.date.accessioned2010-04-09T06:38:54Z-
dc.date.available2010-04-09T06:38:54Z-
dc.date.issued2007-
dc.identifier.citationdental materials 23 (2007) 1307–1312en
dc.identifier.issn0109-5641-
dc.identifier.urihttps://hdl.handle.net/10371/62868-
dc.description.abstractObjectives. The objectives were to measure the difference in the color and color parameters
of natural teeth measured by a tristimulus colorimeter (CM, used as a reference) and
Shade Vision System (SV), and to determine the influence of color parameters on the color
difference between the values measured by two instruments.
Method. Color of 12 maxillary and mandibular anterior teeth was measured by CM and SV
for 47 volunteers (number of teeth = 564). Color parameters such as CIE L*, a* and b* values,
chroma and hue angle measured by two instruments were compared. Chroma was calculated
as C∗ab = (a∗2 + b∗2)1/2, and hue angle was calculated as h◦ = arctan(b*/a*). The influence
of color parameters measured by CM on the color difference ( E∗ab) between the values
measured by two instruments was analyzed with multiple regression analysis (˛ = 0.01).
Results. Mean E∗ab value between the values measured by two instruments was 21.7 (±3.7),
and the mean difference in lightness (CIE L*) and chroma was 16.2 (±3.9) and 13.2 (±3.0),
respectively. Difference in hue angle was high as 132.7 (±53.3)◦. Except for the hue angle, all
the color parameters showed significant correlations and the coefficient of determination
(r2) was in the range of 0.089–0.478. Based on multiple regression analysis, the standardized
partial correlation coefficient (ˇ) of the included predictors for the color difference was
−0.710 for CIE L* and −0.300 for C∗ab (p < 0.01).
Conclusion. All the color parameters showed significant but weak correlations except for hue
angle. When lightness and chroma of teeth were high, color difference between the values
measured by two instruments was small. Clinical accuracy of two instruments should be
investigated further.
en
dc.language.isoenen
dc.publisherElsevieren
dc.subjectTooth coloren
dc.subjectColorimeteren
dc.subjectShade Vision Systemen
dc.subjectColor differenceen
dc.titleComparison of the color of natural teeth measured by a colorimeter and Shade Vision Systemen
dc.typeArticleen
dc.contributor.AlternativeAuthor조병훈-
dc.contributor.AlternativeAuthor임용규-
dc.contributor.AlternativeAuthor이용근-
dc.identifier.doi10.1016/j.dental.2006.11.008-
Appears in Collections:
Files in This Item:
There are no files associated with this item.

Altmetrics

Item View & Download Count

  • mendeley

Items in S-Space are protected by copyright, with all rights reserved, unless otherwise indicated.

Share