Is MRI a reliable tool to locate the electrode after deep brain stimulation surgery? Comparison study of CT and MRI for the localization of electrodes after DBS

Cited 39 time in Web of Science Cited 40 time in Scopus
Authors

Lee, Ji Yeoun; Kim, Jin Wook; Lee, Jee-Young; Lim, Yong Hoon; Kim, Dong Gyu; Paek, Sun Ha; Jeon, Beom Seok; Kim, Cheolyoung

Issue Date
2010-12
Publisher
SPRINGER WIEN
Citation
ACTA NEUROCHIRURGICA; Vol.152 12; 2029-2036
Keywords
CTDBSElectrode localizationParkinson`s diseaseMRIImage fusion
Abstract
MRI has been utilized to localize the electrode after deep brain stimulation, but its accuracy has been questioned due to image distortion. Under the hypothesis that MRI is not adequate for evaluation of electrode position after deep brain stimulation, this study is aimed at validating the accuracy of MRI in electrode localization in comparison with CT scan. Sixty one patients who had undergone STN DBS were enrolled for the analysis. Using mutual information technique, CT and MRI taken at 6 months after the operation were fused. The x and y coordinates of the centers of electrodes shown of CT and MRI were compared in the fused images to calculate average difference at five different levels. The difference of the tips of the electrodes, designated as the z coordinate, was also calculated. The average of the distance between the centers of the electrodes in the five levels estimated in the fused image of brain CT and MRI taken at least 6 months after STN DBS was 1.33 mm (0.1-5.8 mm). The average discrepancy of x coordinates for all five levels between MRI and CT was 0.56 +/- 0.54 mm (0-5.7 mm), the discrepancy of y coordinates was 1.06 +/- 0.59 mm (0-3.5 mm), and for the z coordinate, it was 0.98 +/- 0.52 mm (0-3.1 mm) (all p values < 0.001). Notably, the average discrepancy of x coordinates at 3.5 mm below AC-PC level, i.e., at the STN level between MRI and CT, was 0.59 +/- 0.42 mm (0-2.4 mm); the discrepancy of y coordinates was 0.81 +/- 0.47 mm (0-2.9 mm) (p values < 0.001). The results suggest that there was significant discrepancy between the centers of electrodes estimated by CT and MRI after STN DBS surgery.
ISSN
0001-6268
Language
English
URI
https://hdl.handle.net/10371/77683
DOI
https://doi.org/10.1007/s00701-010-0779-2
Files in This Item:
There are no files associated with this item.
Appears in Collections:

Altmetrics

Item View & Download Count

  • mendeley

Items in S-Space are protected by copyright, with all rights reserved, unless otherwise indicated.

Share