Nonconstrained Sleep Monitoring System and Algorithms Using Air-Mattress With Balancing Tube Method
DC Field | Value | Language |
---|---|---|
dc.contributor.author | Shin, Jae Hyuk | - |
dc.contributor.author | Chee, Young Joon | - |
dc.contributor.author | Jeong, Do-Un | - |
dc.contributor.author | Park, Kwang Suk | - |
dc.date.accessioned | 2012-06-29T02:07:03Z | - |
dc.date.available | 2012-06-29T02:07:03Z | - |
dc.date.issued | 2010-01 | - |
dc.identifier.citation | IEEE TRANSACTIONS ON INFORMATION TECHNOLOGY IN BIOMEDICINE; Vol.14 1; 147-156 | ko_KR |
dc.identifier.issn | 1089-7771 | - |
dc.identifier.uri | https://hdl.handle.net/10371/77881 | - |
dc.description.abstract | We evaluated the performance of a bed-type sensor system using the air-mattress with balancing tube (AMBT) method to noninvasively monitor the signals of heartbeat, respiration, and events of snoring, sleep apnea and body movement of subject on the system. The proposed system consists of multiple cylindrical air cells, two sensor cells and 18 support cells, and the small physiological signals were measured by the changes in pressure difference between the sensor cells, and the dc component was removed by balancing tube that is connecting the sensor cells. Using newly developed AMBT method, heartbeat, respiration, snoring, and body movement signals were clearly measured. For the concept of a home healthcare system, two automatic processing algorithms were developed: one is to estimate the mean heart and respiration rates for every 30 s, and another one is to detect the snoring, sleep apnea, and body movement events from the measured signals. In the beat-to-beat heart rate and breath-by-breath respiration rate analyses, the correlation coefficients of the heart and respiration rates from the proposed AMBT method compared with reference methods, electrocardiogram, and respiration effort signal from piezoelectric belt, were 0.98 (p < 0.01) and 0.96 (p < 0.01), respectively. Sensitivity and positive predictive value (PPV) of the detection algorithm for snoring event were 93%, 96%, for sleep apnea event were 93%, 88%, and for body movement event were 86%, 100%, respectively. These findings support that ABMT method provides an accurate and reliable means to monitor heartbeat, respiration activities and the sleep events during sleep. | ko_KR |
dc.language.iso | en | ko_KR |
dc.publisher | IEEE-INST ELECTRICAL ELECTRONICS ENGINEERS INC | ko_KR |
dc.subject | Air-mattress and balancing tube | ko_KR |
dc.subject | heartbeat | ko_KR |
dc.subject | respiration | ko_KR |
dc.subject | snoring detection | ko_KR |
dc.subject | sleep monitoring | ko_KR |
dc.subject | noninvasive | ko_KR |
dc.title | Nonconstrained Sleep Monitoring System and Algorithms Using Air-Mattress With Balancing Tube Method | ko_KR |
dc.type | Article | ko_KR |
dc.contributor.AlternativeAuthor | 신재혁 | - |
dc.contributor.AlternativeAuthor | 지영준 | - |
dc.contributor.AlternativeAuthor | 정도언 | - |
dc.contributor.AlternativeAuthor | 박광석 | - |
dc.identifier.doi | 10.1109/TITB.2009.2034011 | - |
dc.citation.journaltitle | IEEE TRANSACTIONS ON INFORMATION TECHNOLOGY IN BIOMEDICINE | - |
dc.description.citedreference | Shin JH, 2009, PHYSIOL MEAS, V30, P679, DOI 10.1088/0967-3334/30/7/011 | - |
dc.description.citedreference | Inan OT, 2009, PHYSIOL MEAS, V30, P169, DOI 10.1088/0967-3334/30/2/005 | - |
dc.description.citedreference | Mack DC, 2009, IEEE T INF TECHNOL B, V13, P111, DOI 10.1109/TITB.2008.2007194 | - |
dc.description.citedreference | Han J, 2008, COMPUT METH PROG BIO, V91, P199, DOI 10.1016/j.cmpb.2008.04.012 | - |
dc.description.citedreference | Gonzalez-Landaeta R, 2008, PHYSIOL MEAS, V29, P979, DOI 10.1088/0967-3334/29/8/009 | - |
dc.description.citedreference | Choi BH, 2007, MED BIOL ENG COMPUT, V45, P107, DOI 10.1007/s11517-006-0134-1 | - |
dc.description.citedreference | Zhu X, 2006, IEEE T BIO-MED ENG, V53, P2553, DOI 10.1109/TBME.2006.884641 | - |
dc.description.citedreference | MACK D, 2006, P TRANSD C DISTR DIA, P51 | - |
dc.description.citedreference | Watanabe K, 2005, IEEE T BIO-MED ENG, V52, P2100, DOI 10.1109/TBME.2005.857637 | - |
dc.description.citedreference | Chee Y, 2005, PHYSIOL MEAS, V26, P413, DOI 10.1088/0967-3334/26/4/007 | - |
dc.description.citedreference | Carrington MJ, 2005, J APPL PHYSIOL, V98, P468, DOI 10.1152/japplphysiol.00702.2004 | - |
dc.description.citedreference | Groeger JA, 2004, J SLEEP RES, V13, P359 | - |
dc.description.citedreference | Watanabe T, 2004, IEEE T BIO-MED ENG, V51, P1735, DOI 10.1109/TBME.2004.828037 | - |
dc.description.citedreference | Kundel HL, 2003, RADIOLOGY, V228, P303, DOI 10.1148/radiol.2282011860 | - |
dc.description.citedreference | JANE R, 2003, P 25 ANN INT C IEEE, P366 | - |
dc.description.citedreference | Kohler BU, 2002, IEEE ENG MED BIOL, V21, P42, DOI 10.1109/51.993193 | - |
dc.description.citedreference | Leung RST, 2001, AM J RESP CRIT CARE, V164, P2147, DOI 10.1164/rccm2107045 | - |
dc.description.citedreference | Moore P, 2001, J SLEEP RES, V10, P303 | - |
dc.description.citedreference | Shahar E, 2001, AM J RESP CRIT CARE, V163, P19 | - |
dc.description.citedreference | LAVIE P, 2001, J CLIN HYPERTENS, V3, P296 | - |
dc.description.citedreference | Spiegel K, 1999, LANCET, V354, P1435 | - |
dc.description.citedreference | Rauhala E, 1996, J SLEEP RES, V5, P246 | - |
dc.description.citedreference | Kirjavainen T, 1996, J SLEEP RES, V5, P186 | - |
dc.description.citedreference | HERNANDEZ L, 1995, PHYSIOL MEAS, V16, P161, DOI 10.1088/0967-3334/16/3/003 | - |
dc.description.citedreference | SOMERS VK, 1993, NEW ENGL J MED, V328, P303 | - |
dc.description.citedreference | JANSEN BH, 1991, IEEE T BIO-MED ENG, V38, P748, DOI 10.1109/10.83586 | - |
dc.description.citedreference | HUNG J, 1990, LANCET, V336, P261 | - |
dc.description.citedreference | BLAND JM, 1986, LANCET, V1, P307 | - |
dc.description.citedreference | CARSKADON MA, 1980, SLEEP, V2, P309 | - |
dc.description.citedreference | RECHTSCHAFFEN A, 1977, MANUAL STANDARDIZED | - |
dc.description.citedreference | DEUCHAR DC, 1967, BRIT HEART J, V29, P285 | - |
dc.description.citedreference | POLLOCK P, 1957, CAN MED ASSOC J, V76, P778 | - |
dc.description.citedreference | STARR I, 1950, CIRCULATION, V1, P1073 | - |
dc.description.tc | 8 | - |
- Appears in Collections:
- Files in This Item:
- There are no files associated with this item.
Item View & Download Count
Items in S-Space are protected by copyright, with all rights reserved, unless otherwise indicated.