Publications
Detailed Information
Ceramic fracture following cervical disc arthroplasty
DC Field | Value | Language |
---|---|---|
dc.contributor.author | Nguyen, Ngoc Quyen | - |
dc.contributor.author | Kafle, Dinesh | - |
dc.contributor.author | Buchowski, Jacob M. | - |
dc.contributor.author | Park, Kun-Woo | - |
dc.contributor.author | Chang, Bong-Soon | - |
dc.contributor.author | Lee, Choon-Ki | - |
dc.contributor.author | Yeom, Jin S. | - |
dc.creator | 염진섭 | - |
dc.date.accessioned | 2013-01-25T06:24:24Z | - |
dc.date.available | 2013-01-25T06:24:24Z | - |
dc.date.issued | 2011-11 | - |
dc.identifier.citation | JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME Vol.93 No.e132, pp. 1-4 | - |
dc.identifier.issn | 0021-9355 | - |
dc.identifier.uri | https://hdl.handle.net/10371/81094 | - |
dc.description.abstract | Although ceramic fractures have been reported following
total hip arthroplasty1-6, with an incidence ranging from 0.004% to 1.4%1-3, no cases of ceramic fracture following cervical disc arthroplasty have been reported, to our knowledge. We present the case of a patient with cervical radiculopathy who underwent total disc replacement at C5-C6 and C6-C7, which was complicated by the development of recurrent symptoms approximately one month after the index procedure. At the time of the revision surgery, the ceramic insert at the C6-C7 level was found to be fractured on its convex cranial side. The implant used was a semiconstrained ceramicon- ceramic prosthesis (Discocerv Cervidisc Evolution; Scientx, Guyancourt, France). Since the introduction of this prosthesis in April 20067, more than 2000 prostheses have been implanted. The patient was informed that data concerning the case would be submitted for publication, and he gave his consent. | - |
dc.description.uri | Copyright(2011) THE JOURNAL OF BONE AND JOINT SURGERY, INCORPORATED. This article may be downloaded for personal use only. Any other use requires prior permission of the author and the THE JOURNAL OF BONE AND JOINT SURGERY, INCORPORATED.
The following article appeared in (citation of published article) and may be found at related URL. | - |
dc.language.iso | en | en |
dc.publisher | JOURNAL BONE JOINT SURGERY INC | en |
dc.subject | 복합학 | en |
dc.title | Ceramic fracture following cervical disc arthroplasty | en |
dc.type | Article | - |
dc.contributor.AlternativeAuthor | 박근우 | - |
dc.contributor.AlternativeAuthor | 장봉순 | - |
dc.contributor.AlternativeAuthor | 이준기 | - |
dc.contributor.AlternativeAuthor | 염진섭 | - |
dc.identifier.doi | 10.2106/JBJS.K.00077 | - |
dc.description.srnd | OAIID:oai:osos.snu.ac.kr:snu2011-01/102/0000004226/8 | - |
dc.description.srnd | SEQ:8 | - |
dc.description.srnd | PERF_CD:SNU2011-01 | - |
dc.description.srnd | EVAL_ITEM_CD:102 | - |
dc.description.srnd | USER_ID:0000004226 | - |
dc.description.srnd | ADJUST_YN:N | - |
dc.description.srnd | EMP_ID:A076317 | - |
dc.description.srnd | DEPT_CD:801 | - |
dc.description.srnd | CITE_RATE:3.272 | - |
dc.description.srnd | FILENAME:E036T_JBJS-A-2011_Nguyen_Ceramic fracture following cervical disc arthroplasty.pdf | - |
dc.description.srnd | DEPT_NM:의학과 | - |
dc.description.srnd | EMAIL:highcervical@gmail.com | - |
dc.description.srnd | SCOPUS_YN:N | - |
dc.description.srnd | CONFIRM:Y | - |
dc.identifier.srnd | 2011-01/102/0000004226/8 | - |
- Appears in Collections:
Item View & Download Count
Items in S-Space are protected by copyright, with all rights reserved, unless otherwise indicated.