Publications

Detailed Information

Modification of glassy carbon electrode with poly-melamine film: An application for the detection of hydroquinone, catechol and resorcinol : 폴리-멜라민 필름을 이용한 유리탄소전극의 변형: 하이드로퀴논, 카테콜, 레조시놀 검출에 적용

DC Field Value Language
dc.contributor.advisor홍훈기-
dc.contributor.author김은경-
dc.date.accessioned2017-07-19T02:58:08Z-
dc.date.available2017-07-19T02:58:08Z-
dc.date.issued2015-02-
dc.identifier.other000000025123-
dc.identifier.urihttps://hdl.handle.net/10371/128039-
dc.description학위논문 (석사)-- 서울대학교 대학원 : 과학교육과, 2015. 2. 홍훈기.-
dc.description.abstractIn this study, we share findings about the development of a sensor using poly-melamine modified glassy carbon electrode (PMel/GCE) that is capable of detecting dihydroxybenzene isomers simultaneously in aqueous solution. We prepared poly-melamine film by electro-polymerization on GCE in acidic solution containing melamine monomer, then it was characterized by field emission scanning electron microscopy (FE-SEM), electrochemical impedance spectroscopy (EIS). Cyclic voltammetry (CV) shows the electrochemical behaviors of dihydroxybenzene isomers at the PMel/GCE. Square wave voltammetry (SWV) was used for the simultaneous determination of dihydroxybenzene isomers. The two broad anodic peaks of dihydroxybenzene isomers at 0.380 V for hydroquinone and catechol and 0.712 V for resorcinol on the bare GCE resolved in three well-defined peaks at 0.084 V for hydroquinone, 0.184 V for catechol and 0.608 V for resorcinol vs. Ag/AgCl, respectively. It exhibited a good linearity to dihydroxybenzene isomers in the concentration range of 5-300 µM for hydroquinone, 5-500 µM for catechol and 3-200 µM for resorcinol with good sensitivity of 0.052, 0.055 and 0.038 µA/µM. And the detection limits were 0.65, 1.90 and 0.58 µM at a signal-to-noise ratio of 3. Furthermore, this sensor is expected to be utilized in detecting these isomers in local tap water sample with good anti-interference.-
dc.description.tableofcontents1. Introduction 2

2. Experimental 5

2.1. Apparatus and reagents 5
2.2. Preparation of the PMel/GCE 6
2.3. Electrochemical measurements 6

3. Results and Discussion 8

3.1. Fabrication of the Poly-melamine film 8
3.2. Characterization of the Poly-melamine film 11
3.3. Electrochemical behaviors of HQ, CA and RE 15
3.4. Effect of different scan rate 18
3.5. Effect of buffer pH 20
3.6. Simultaneous determination of HQ, CA and RE 23
3.7. Stability and Reproducibility 28
3.8. Sample analysis 29

4. Conclusion 31

5. References 32
-
dc.formatapplication/pdf-
dc.format.extent924877 bytes-
dc.format.mediumapplication/pdf-
dc.language.isoen-
dc.publisher서울대학교 대학원-
dc.subject멜라민-
dc.subject.ddc507-
dc.titleModification of glassy carbon electrode with poly-melamine film: An application for the detection of hydroquinone, catechol and resorcinol-
dc.title.alternative폴리-멜라민 필름을 이용한 유리탄소전극의 변형: 하이드로퀴논, 카테콜, 레조시놀 검출에 적용-
dc.typeThesis-
dc.contributor.AlternativeAuthorEun-Kyoung Kim-
dc.description.degreeMaster-
dc.citation.pages36-
dc.contributor.affiliation사범대학 과학교육과-
dc.date.awarded2015-02-
Appears in Collections:
Files in This Item:

Altmetrics

Item View & Download Count

  • mendeley

Items in S-Space are protected by copyright, with all rights reserved, unless otherwise indicated.

Share