Removal Efficiency of Volatile Organic Compounds using Different Types of Activated Charcoal in the Fume Hood : 흄후드에서 활성탄 형태에 따른 휘발성 유기화합물 제거에 관한 연구

DC Field Value Language
dc.contributor.advisor윤충식-
dc.contributor.author이상아-
dc.date.accessioned2017-07-19T03:10:29Z-
dc.date.available2017-07-19T03:10:29Z-
dc.date.issued2015-08-
dc.identifier.other000000066885-
dc.identifier.urihttps://hdl.handle.net/10371/128233-
dc.description학위논문 (석사)-- 서울대학교 보건대학원 : 환경보건학과(산업보건전공), 2015. 8. 윤충식.-
dc.description.abstractIn laboratories, fume hoods are used to handle most chemicals, with the goal of protecting the health of workers and preventing the deterioration of indoor air quality. However, there is no obligation to install air purifiers in fume hood systems in Korea. As a result, harmful vapors and gases form fume hoods can be emitted directly into the atmosphere through the exhaust duct. This study compared the breakthrough time of three different types of activated charcoal adsorbent, and evaluated their capacity to adsorb volatile organic compounds (VOCs) within a local exhaust ventilation (LEV) system.
Acetone was used as a surrogate for polar volatile organic compounds (VOCs) and toluene was used as a surrogate for non-polar VOCs. Both were tested against three different kinds of adsorbent: an activated carbon fiber (ACF), granular activated carbon (GAC), and an ACF + GAC mixture.
Approximately 10 and 50 ppm of the tested solvents were evaporated by a syringe pump in a hood, and the breakthrough time was determined by comparing the concentrations measured by a real-time monitor before and after adsorption. The adsorbents were installed 5 m from the hood as part of the LEV system.
The breakthrough time for toluene was on average two times longer (>1.1 - 2.8 times) than for acetone for all three adsorbents. When the 50 ppm concentration was used, the breakthrough time was 3.1 times shorter (>1.3 - 4.4) than when the 10 ppm concentration was used. The ACF + GAC mixture had the longest breakthrough time for both acetone and toluene regardless of the concentration used, followed by GAC and then ACF. The breakthrough time was reduced by 7.4 - 9.1 times when the length of the ACF + GAC was shortened from 14 to 7 cm. The adsorption efficiency for acetone and toluene was 75.6 and 80.7% of the total evaporated amount before breakthrough, respectively. The ACF + GAC mixture was proven to have a long breakthrough time and the adsorption rate was relatively stable until breakthrough.
Among the three types of adsorbent the ACF + GAC mixture was found to have the longest breakthrough time for both toluene and acetone. An ACF + GAC mixture is therefore considered suitable for use within a laboratory LEV system to improve air quality.
-
dc.description.tableofcontentsABSTRACT 1
Contents 4
List of Tables 5
List of Figures 6
1. Introduction 7
2. Materials and method 9
2.1. Study design 9
2.2. Activated Carbon Adsorption 11
2.2.1. VOCs vaporization 13
2.3. Sampling method 15
2.4. Calculation and statistics 17
3. Results 18
3.1. Comparison of adsorbents 18
3.2. Comparison of adsorption length 23
3.3. Comparison of the breakthrough curve 25
3.3.1. Breakthrough curve for acetone 25
3.3.2. Breakthrough curve for toluene 27
3.4. Adsorption capacity 29
4. Discussion 33
5. Conclusions 36
6. References 37
-
dc.formatapplication/pdf-
dc.format.extent1549001 bytes-
dc.format.mediumapplication/pdf-
dc.language.isoen-
dc.publisher서울대학교 보건대학원-
dc.subjectFume hood-
dc.subjectActivated Carbon-
dc.subjectVOCs-
dc.subject.ddc363-
dc.titleRemoval Efficiency of Volatile Organic Compounds using Different Types of Activated Charcoal in the Fume Hood-
dc.title.alternative흄후드에서 활성탄 형태에 따른 휘발성 유기화합물 제거에 관한 연구-
dc.typeThesis-
dc.contributor.AlternativeAuthorSang ah Lee-
dc.description.degreeMaster-
dc.citation.pages47-
dc.contributor.affiliation보건대학원 환경보건학과-
dc.date.awarded2015-08-
Appears in Collections:
Files in This Item:

Altmetrics

Item View & Download Count

  • mendeley

Items in S-Space are protected by copyright, with all rights reserved, unless otherwise indicated.

Share