Devising negative pressure within intercuff space reduces microaspiration

DC Field Value Language
dc.contributor.authorSohn, H. M.-
dc.contributor.authorBaik, J. S.-
dc.contributor.authorHwang, J. Y.-
dc.contributor.authorKim, S. Y.-
dc.contributor.authorHan, S. H.-
dc.contributor.authorKim, J. H.-
dc.date.accessioned2019-03-12T00:29:40Z-
dc.date.available2019-03-12T09:30:37Z-
dc.date.issued2018-12-03-
dc.identifier.citationBMC Anesthesiology, 18(1):181ko_KR
dc.identifier.issn1471-2253-
dc.identifier.urihttps://hdl.handle.net/10371/146962-
dc.description.abstractBackground
Microaspiration past the tracheal tube cuffs causes ventilator-associated pneumonia. The objective of the current study was to evaluate whether creating negative pressure between the tracheal double cuffs could block the fluid passage past the tracheal tube cuffs.

Methods
A new negative pressure system was devised between the double cuffs through a suction hole in the intercuff space. Blue-dyed water was instilled above the cuff at negative suction pressures of − 54, − 68, − 82, − 95, − 109, − 122, and − 136 cmH2O, and the volume leaked was measured in an underlying water trap after 10 min. Leakage tests were also performed during positive pressure ventilation, and using higher-viscosity materials. The actual negative pressures delivered at the hole of double cuffs were obtained by placing microcatheter tip between the intercuff space and the artificial trachea.

Results
No leakage occurred past the double cuff at − 136 cmH2O suction pressure at all tracheal tube cuff pressures. The volume leaked decreased significantly as suction pressure increased. When connected to a mechanical ventilator, no leakage was found at − 54 cmH2suction pressure. Volume of the higher-viscosity materials (dynamic viscosity of 63–108 cP and 370–430 cP) leaked was small compared to that of normal saline (0.9–1.1 cP). The pressures measured in the intercuff space corresponded to 3.8–5.9% of those applied.

Conclusions
A new prototype double cuff with negative pressure in the intercuff space completely prevented water leakage. The negative pressure transmitted to the tracheal inner wall was a small percentage of that applied.
ko_KR
dc.description.sponsorshipThis work was supported by grant 02–2013-065 from the SNUBH Research Fundko_KR
dc.language.isoenko_KR
dc.publisherBioMed Centralko_KR
dc.subjectArtificial tracheako_KR
dc.subjectAspirationko_KR
dc.subjectDouble cuffsko_KR
dc.subjectEndotracheal tubeko_KR
dc.subjectNegative pressureko_KR
dc.titleDevising negative pressure within intercuff space reduces microaspirationko_KR
dc.typeArticleko_KR
dc.identifier.doi10.1186/s12871-018-0643-0-
dc.language.rfc3066en-
dc.rights.holderThe Author(s).-
dc.date.updated2018-12-09T04:25:29Z-
Appears in Collections:
Files in This Item:

Altmetrics

Item View & Download Count

  • mendeley

Items in S-Space are protected by copyright, with all rights reserved, unless otherwise indicated.

Share