Indirect Clinical Validation of a Programmed Death-Ligand 1 Laboratory-Developed Test for Gastric/Gastroesophageal Junction Adenocarcinoma with 22C3 Antibody Concentrate

Cited 2 time in Web of Science Cited 3 time in Scopus
Authors

Kim, Ji Min; Kim, Binnari; Kim, Eunji; Jang, Minsun; Cho, Jun Hun; Lee, Hye Seung; Kwak, Yoonjin; Huang, Lingkang; Krishnan, Radha; Bai, Sally Y.; Mounawar, Mounia; Kim, Kyoung-Mee

Issue Date
2022-11
Publisher
Adis International Ltd.
Citation
Molecular Diagnosis and Therapy, Vol.26 No.6, pp.679-688
Abstract
Background The PD-L1 IHC 22C3 pharmDx used on the Dako Autostainer Link 48 (ASL48) staining platform is an established method for assessing programmed death-ligand 1 (PD-L1) expression in tumor tissue and determining patient eligibility for pembrolizumab treatment; however, the availability of this platform is limited in Europe and Asia. Objectives The aims of this study were to develop and optimize protocols for the PD-L1 22C3 antibody concentrate with multiple immunohistochemistry staining platforms and to validate these protocols using PD-L1 combined positive score (CPS) with a cut-off of >= 1 in gastric or gastroesophageal junction adenocarcinoma. Design The 22C3 antibody concentrate was tested and optimized protocols were developed for use with three staining platforms: Dako ASL48, Ventana BenchMark ULTRA, and Leica BOND-MAX. Tumor specimens (N = 120) from patients with gastric or gastroesophageal junction adenocarcinoma were used for the validation study; these specimens were evaluated independently by three pathologists for PD-L1 CPS as a continuous variable and using a cut-off of >= 1. PD-L1 IHC 22C3 pharmDx used on the Dako ASL48 platform served as the reference or gold standard. Results The intraclass correlation coefficient of CPS as a continuous variable between the gold standard and each staining platform assessed was 0.910-0.989. When CPS was dichotomized based on a cut-off of >= 1, depending on the pathologist and the platform used, positive percentage agreement was 81-99% and negative percentage agreement was 90-100%. Interobserver agreement using the gold standard showed substantial agreement (kappa = 0.779). Conclusion The PD-L1 22C3 antibody concentrate can potentially be used with the laboratory-developed test on three commercially available immunohistochemistry staining platforms to determine PD-L1 expression in tumor samples from patients with gastric or gastroesophageal junction adenocarcinoma.
ISSN
1177-1062
URI
https://hdl.handle.net/10371/188896
DOI
https://doi.org/10.1007/s40291-022-00605-2
Files in This Item:
There are no files associated with this item.
Appears in Collections:

Altmetrics

Item View & Download Count

  • mendeley

Items in S-Space are protected by copyright, with all rights reserved, unless otherwise indicated.

Share