Comparison of the International Normalized Ratio Between a Point-of-Care Test and a Conventional Laboratory Test: the Latter Performs Better in Assessing Warfarin-induced Changes in Coagulation Factors
Cited 2 time in
Web of Science
Cited 4 time in Scopus
- Authors
- Issue Date
- 2023-07
- Publisher
- 대한진단검사의학회
- Citation
- Annals of Laboratory Medicine, Vol.43 No.4, pp.337-344
- Abstract
- Background: Point-of-care testing (POCT) coagulometers are increasingly used for monitoring warfarin therapy. However, in high international normalized ratio (INR) ranges, significant discrepancy in the INR between POCT and conventional laboratory tests occurs. We compared the INR of POCT (CoaguChek XS Plus; Roche Diagnostics, Mannheim, Germany) with that of a conventional laboratory test (ACL TOP 750; Instrumentation Laboratory SpA, Milan, Italy) and explored possible reasons for discrepancy. Methods: Paired POCT and conventional laboratory test INRs were analyzed in 400 samples from 126 patients undergoing warfarin therapy after cardiac surgery. Coagulation factor and thrombin generation tests were compared using the Mann-Whitney U test. Correlations between coagulation factors and INRs were determined using Pearson correlation coefficients. Results: The mean difference in the INR between the tests increased at high INR ranges. Endogenous thrombin potential levels were decreased at INR <2.0 for CoaguChek XS Plus and 2.0< INR <3.0 for ACL TOP 750 compared with those at INR <2.0 for both tests, indicating a better performance of ACL TOP 750 in assessing thrombin changes. The correlation coefficients of coagulation factors were stronger for ACL TOP 750 INR than for CoaguChek XS Plus INR. Vitamin K-dependent coagulation factors were found to contribute to the INR discrepancy. Conclusions: Decreases in vitamin K-dependent coagulation and anticoagulation factors can explain the significant discrepancy between the two tests in high INR ranges. Since conventional laboratory test INR values are more reliable than POCT INR values, a confirmatory conventional laboratory test is required for high INR ranges.
- ISSN
- 2234-3806
- Files in This Item:
- There are no files associated with this item.
- Appears in Collections:
Item View & Download Count
Items in S-Space are protected by copyright, with all rights reserved, unless otherwise indicated.