Publications

Detailed Information

Novel robotic catheter remote control system: feasibility and safety of transseptal puncture and endocardial catheter navigation

DC Field Value Language
dc.contributor.authorSaliba, Walid-
dc.contributor.authorCummings, Jennifer E-
dc.contributor.authorOh, Seil-
dc.contributor.authorZhang, Youhua-
dc.contributor.authorMazgalev, Todor N-
dc.contributor.authorSchweikert, Robert A-
dc.contributor.authorBurkhardt, J David-
dc.contributor.authorNatale, Andrea-
dc.date.accessioned2010-01-25T12:47:06Z-
dc.date.available2010-01-25T12:47:06Z-
dc.date.issued2006-08-02-
dc.identifier.citationJ Cardiovasc Electrophysiol. 2006 Oct;17(10):1102-5. Epub 2006 Jul 18.en
dc.identifier.issn1540-8167 (Electronic)-
dc.identifier.urihttp://www.ncbi.nlm.nih.gov/entrez/query.fcgi?cmd=Retrieve&db=PubMed&dopt=Citation&list_uids=16879628-
dc.identifier.urihttps://hdl.handle.net/10371/44042-
dc.description.abstractOBJECTIVES: The aims of this study were to demonstrate the safety and the feasibility of the robotic catheter remote control system (CCS) in endocardial navigation in all cardiac chambers, as well as facilitation of the transseptal puncture. BACKGROUND: CCS has been developed to facilitate control and precise positioning of catheters within the cardiovascular system. METHODS: CCS consists of a remote catheter manipulator, a set up joint, a physician workstation, and a steerable guide catheter (SGC) and sheath. A conventional 4-mm tip catheter was inserted through the SGC to perform mapping of five predefined targets in each cardiac chamber. Seven mongrel dogs were used in this study. Intracardiac echocardiography and three-dimensional (3-D) electroanatomical mapping were integrated with CCS to facilitate catheter manipulation and to guide transseptal puncture. The time to complete the transseptal puncture and the time to complete access to the predefined targets in each cardiac chamber were measured. Gross and microscopic examinations of the accessed and ablation sites were performed to evaluate safety. RESULTS: Transseptal puncture was performed successfully in all animals with a mean time of 7 +/- 3 minutes. Procedure times to access the five targets in the right atrium, right ventricle, left atrium, and left ventricle were 5.6 +/- 1.7, 4.6 +/- 1.5, 13.5 +/- 11.0, 7.0 +/- 2.9 minutes, respectively. There were no intracardiac damages associated with catheter manipulation noted in the excised hearts. CONCLUSIONS: Endocardial catheter navigation and mapping using the robotic catheter remote control is safe and feasible. Moreover, the CCS could be used to perform transseptal puncture and left atrial instrumentation.en
dc.language.isoenen
dc.publisherBlackwell Publishingen
dc.subjectAnimalsen
dc.subjectCatheter Ablation/adverse effects/*instrumentation/methodsen
dc.subjectDogsen
dc.subjectEndocardium/surgery/ultrasonographyen
dc.subjectEquipment Designen
dc.subjectEquipment Failure Analysisen
dc.subjectEquipment Safetyen
dc.subjectFeasibility Studiesen
dc.subjectHeart Catheterization/adverse effects/*instrumentation/methodsen
dc.subjectHeart Injuries/etiology/ultrasonographyen
dc.subjectHeart Septum/surgery/ultrasonographyen
dc.subjectPilot Projectsen
dc.subjectPunctures/adverse effects/*instrumentation/methodsen
dc.subjectRobotics/*instrumentation/methodsen
dc.subjectSurgery, Computer-Assisted/*instrumentation/methodsen
dc.subjectTelemedicine/*instrumentation/methodsen
dc.titleNovel robotic catheter remote control system: feasibility and safety of transseptal puncture and endocardial catheter navigationen
dc.typeArticleen
dc.contributor.AlternativeAuthor오세일-
dc.identifier.doi10.1111/j.1540-8167.2006.00556.x-
Appears in Collections: