Browse
S-Space
College of Dentistry/School of Dentistry (치과대학/치의학대학원)
Dept. of Dentistry (치의학과)
Journal Papers (저널논문_치의학과)
임플랜트 상부구조의 재료가 반복하중 후 나사풀림에 미치는 영향 : AN INFLUENCE OF ABUTMENT MATERIALS ON A SCREW-LOOSENING AFTER CYCLIC LOADING
DC Field | Value | Language |
---|---|---|
dc.contributor.author | 이태식 | - |
dc.contributor.author | 한중석 | - |
dc.contributor.author | 양재호 | - |
dc.contributor.author | 이재봉 | - |
dc.contributor.author | 김성훈 | - |
dc.date.accessioned | 2010-02-01T06:14:55Z | - |
dc.date.available | 2010-02-01T06:14:55Z | - |
dc.date.issued | 2007 | - |
dc.identifier.citation | 2007;45(2) | en |
dc.identifier.issn | 0301-2875 | - |
dc.identifier.uri | https://hdl.handle.net/10371/47494 | - |
dc.description.abstract | Statement of problem: A phenomenon of screw-loosening in implant abutment is frequently occurred in a single and multiple implant restoration. Purpose: This study was performed to evaluate an effect of abutment material on screw-loosening before and after a cyclic loading. In a single-tooth implant, different materials of abutment, Type III Gold alloy and Zirconium composite were used. Material and method: The Gold alloy(Type III) and Zirconium composite were used to make a superstructure of implant, the one of types of UCLA, Each group was constituted of 5 sample with a 30-degree offset angulated loading platform. The external hexagonal fixture was rigidly hel d in a special holding zig to ensure solid fixation without rotation during the tightening and a cyclic loading. A Titanium-alloy screw was used to connect and controlled to be tighten in 20Ncm torque by a digital torque gauge. A 20 times of consecutive closing/opening cycle were performed to evaluate the immediate torque loss. In 5 sample of each material group, an initial opening torque was recorded during 3 closing/opening cycle, then 2Hz, 200N, 1,000,000 cyclic loadings were performed, then a opening torque was evaluated. Result & Conclusion: 1. In this limited study, titanium alloy screw tightened in 20Ncm, a cold-welding phenomen on was not observed during the 20 times of closing/opening cycle(p=0.11, p=0.18). 2. In titanium alloy abutment screw, repeated opening and closing of the screw caused to progressive decrease of opening torque(p=0.014). 3. The difference in preload of screw between gold alloy abutment and ceramic abutment was not significant(p=0.78). 4. The difference in torque loss of screw between gold alloy abutment and ceramic abutment was not significant after 2Hz,200N, 1,000,000 cyclic loading(p=0.92). 5. In titanium alloy abutment screw tightened by 20Ncm, the screw loosening was not significant on each group after 2Hz, 200Ncm, 1,000,000 cyclic loading(p=0.59). | en |
dc.language.iso | ko | en |
dc.publisher | 대한치과보철학회 | en |
dc.subject | Implant | en |
dc.subject | Cyclic loading | en |
dc.subject | Screw loosening | en |
dc.title | 임플랜트 상부구조의 재료가 반복하중 후 나사풀림에 미치는 영향 | en |
dc.title.alternative | AN INFLUENCE OF ABUTMENT MATERIALS ON A SCREW-LOOSENING AFTER CYCLIC LOADING | en |
dc.type | Article | en |
dc.contributor.AlternativeAuthor | Lee, Tae-Sik | - |
dc.contributor.AlternativeAuthor | Han, Jung-Suk | - |
dc.contributor.AlternativeAuthor | Yang, Jae-Ho | - |
dc.contributor.AlternativeAuthor | Lee, Jae-Bong | - |
dc.contributor.AlternativeAuthor | Kim, Sung-Hun | - |
Items in S-Space are protected by copyright, with all rights reserved, unless otherwise indicated.