Publications
Detailed Information
Resonance Frequency Analysis of Implants with Anodized Surface Oxides
DC Field | Value | Language |
---|---|---|
dc.contributor.author | Choi, Jeong-Won | - |
dc.contributor.author | Heo, Seong-Joo | - |
dc.contributor.author | Chang, Ik-Tae | - |
dc.contributor.author | Koak, Jai-Young | - |
dc.contributor.author | Han, Jong-Hyun | - |
dc.contributor.author | Kim, Yong-Sik | - |
dc.contributor.author | Lee, Seok-Hyung | - |
dc.contributor.author | Yim, Soon-Ho | - |
dc.date.accessioned | 2010-02-01T09:57:15Z | - |
dc.date.available | 2010-02-01T09:57:15Z | - |
dc.date.issued | 2004 | - |
dc.identifier.citation | 대한치과보철학회지 2004;42: 294-300. | en |
dc.identifier.issn | 0301-2875 | - |
dc.identifier.uri | http://uci.or.kr/ | - |
dc.identifier.uri | https://hdl.handle.net/10371/47915 | - |
dc.description.abstract | The present experimental study was designed to address two issues. The first was to investigate whether oxidation voltage of titanium implants influenced bone tissue responses after an in vivo implantation. The second aim was to investigate secondary stability change after 1 to 3months period. Screw-shaped implants with a wide range of oxide properties were prepared by electrochemical oxidation methods, where the oxide thickness varied in the range of . The micro structure revealed pore sizes of , the crystal structures of the titanium oxide were amorphous, anatase and a mixture of anatase and rutile type. Bone tissue responses were evaluated by resonance frequency measurements that were undertaken 1 to 3months after insertion in the rabbit tibia. It was concluded that no statistical difference of RFA values was found between the groups, RFA gains after Imonth and 3months were calculated. | en |
dc.description.sponsorship | This study was supported by a grant of the Korea Health 21 R & D Project. Ministry of Health & Welfare. Republic of Korea | en |
dc.language.iso | en | en |
dc.publisher | 대한치과보철학회 | en |
dc.subject | Anodic oxidation | en |
dc.subject | Resonance frequency analysis | en |
dc.subject | Titanium implant | en |
dc.subject | Voltage | en |
dc.title | Resonance Frequency Analysis of Implants with Anodized Surface Oxides | en |
dc.type | Article | en |
dc.contributor.AlternativeAuthor | 최정원 | - |
dc.contributor.AlternativeAuthor | 허성주 | - |
dc.contributor.AlternativeAuthor | 장익태 | - |
dc.contributor.AlternativeAuthor | 곽재영 | - |
dc.contributor.AlternativeAuthor | 한종현 | - |
dc.contributor.AlternativeAuthor | 김용식 | - |
dc.contributor.AlternativeAuthor | 이석형 | - |
dc.contributor.AlternativeAuthor | 임순호 | - |
- Appears in Collections:
Item View & Download Count
Items in S-Space are protected by copyright, with all rights reserved, unless otherwise indicated.