Cited 15 time in Web of Science Cited 16 time in Scopus
Authors

Heo, Min-Suk; Choi, Da-Hye; Benavides, Erika; Huh, Kyung-Hoe; Yi, Won-Jin; Lee, Sam-Sun; Choi, Soon-Chul

Issue Date
2009-08
Publisher
Mosby
Citation
Oral Surg Oral Med Oral Pathol Oral Radiol Endod 2009;108:278-283
Abstract
Objectives
The purpose of this study is to investigate the effects of different bit depths and kilovoltage peak (kVp) values used in intraoral digital radiography on observer performance in detecting subtle radiographic density differences.

Study design
Using an intraoral CCD sensor set at 8- or 12-bit depth, kVp set at 60 or 70, and 14 different exposure times, digital radiographs were acquired of a specially designed aluminum step-wedge phantom with small holes of different depths in each step. Ten observers examined all images.

Results
The observers counted more holes with the 12-bit images compared with the 8-bit images, particularly at 60 kVp. Significantly more holes were also counted with images taken at 70 kVp compared with those taken at 60 kVp.

Conclusion
Regarding the detection of subtle radiographic density differences, 12-bit images and 70 kVp were superior to 8-bit images and 60 kVp
ISSN
1079-2104
Language
English
URI
https://hdl.handle.net/10371/62164
DOI
https://doi.org/10.1016/j.tripleo.2008.12.053
Files in This Item:
There are no files associated with this item.
Appears in Collections:

Altmetrics

Item View & Download Count

  • mendeley

Items in S-Space are protected by copyright, with all rights reserved, unless otherwise indicated.

Share