파노라마방사선촬영기종에 따른 흡수선량 차이 : Difference in Radiation Absorbed Dose According to the Panoramic Radiographic Machines
- Authors
- Issue Date
- 2000-03
- Publisher
- 대한구강악안면방사선학회
- Citation
- Korean J Oral Maxillofac Radiol 2000;30:11-15
- Keywords
- absorbed dose ; dosimetry ; TLD ; panoramic radiography
- Abstract
- Purpose: The primary objective of this study was to estimate the radiation absorbed doses in certain critical organs in the head and neck region with an Orthopos plus, a Panelipse, and a Panex-E machines. The second objective was to compare the absorbed doses between 5 inch by 12 inch and 6 inch by 12 inch image field for the Orthopos plus. Materials and Methods: Rando phantom and LiF TLD chips were used for dosimetry. The absorbed doses were measured at the thyroid gland, the submandibular gland, the parotid gland, the mouth floor, the maxillary sinus, the brain, the mandibular body, the mandibular ramus, the 2nd cervical spine and the skin over TMJ area. Results: The highest absorbed dose value was recorded at the mandibular ramus for the Orthopos plus with narrow image field. Higher absorbed dose values were recorded at the parotid gland, the mouth floor, the submandibular gland, and the 2nd cervical spine. The doses in the parotid gland were 597 μGy and 529 μGy with Orthopos plus, 638 μGy with Panelipse, and 1094μGy with Panex-E. Corresponding figures for the mandibular ramus were 2363 Gy and 1220 μGy, 248 μGy, and 118 μGy. The absorbed doses to the thyroid gland, the maxillary sinus, the brain, and the skin over TMJ were very low. Conclusion: Higher exposure values were recorded for the Orthopos plus than Panelipse and Panex-E. There was no significant differences of the absorbed doses according to the image field size.
- ISSN
- 1229-8212
- Language
- Korean
- Files in This Item:
Items in S-Space are protected by copyright, with all rights reserved, unless otherwise indicated.