A strain gauge analysis of implant-supported cantilevered fixed prosthesis under distal static load
DC Field | Value | Language |
---|---|---|
dc.contributor.author | Sohn, Byoung-Sup | - |
dc.contributor.author | Heo, Seong-Joo | - |
dc.contributor.author | Chang, Ik-Tae | - |
dc.contributor.author | Koak, Jai-Young | - |
dc.contributor.author | Kim, Seong-Kyun | - |
dc.date.accessioned | 2011-10-18 | - |
dc.date.available | 2011-10-18 | - |
dc.date.issued | 2007-12 | - |
dc.identifier.citation | 대한치과보철학회지 45:717-723,2007 | en |
dc.identifier.issn | 0301-2875 | - |
dc.identifier.uri | https://hdl.handle.net/10371/74241 | - |
dc.description.abstract | Statement of problem. Unreasonable distal cantilevered implant-supported prosthesis
can mask functional problems of reconstruction temporarily, but it can cause serious strain and stress around its supported implant and surrounding alveolar bone. Purpose. The purpose of this study was to evaluate strain of implants supporting distal cantilevered fixed prosthesis with two different cantilevered length under distal cantilevered static load. Material and methods. A partially edentulous mandibular test model was fabricated with auto-polymerizing resin (POLYUROCK; Metalor technologies, Stuttgart, Swiss) and artificial denture teeth (Endura; Shoh inc., Kyoto, Japan). Two implants-supported 5-unit screwretained cantilevered fixed prosthesis was made using standard methods with Type Ill gold alloy (Harmony C&B55; Ivoclar-vivadent, Liechtenstein, Germany) for superstructure and reinforced hard resin (Tescera; Ivoclar-vivadent, Liechtenstein, Germany) for occlusal material. Two strain gauges (KFG-1-120-C1-1lLlM2R; KYOWA electronic instruments, Tokyo, Japan) were then attached to the mesial and the distal surface of each standard abutment with adhesive (M-bond 200; Tokuyama, Tokyo, Japan). Total four strain gauges were attached to test model and connected to dynamic signal conditioning strain amplifier (CTA1000; Curiotech inc., Paju, Korea). The stepped 20-100 N in 25 N increments, cantilevered static load 8rnm apart (Group I ) or 16mm apart (Group 11 ), were applied using digital push-pull gauge (Push-Pull Scale & Digital Force Gauge, Axis inc., Seoul, Korea). Each step was performed ten times and every strain signal was monitored and recorded. Results. In case of Group I, the strain values were surveyed by 80.7 - 353.8 pm /m in Chl, 7.5 - 47.9 p /m in CQ45.7 - 278.6 rn / m in Ch3 and -212.2 - -718.7 rn / m in Ch4 depending on increasing cantilevered static load. On the other hand, the strain values of Group II were surveyed by 149.9 - 612.8 0 /m in Chl, 26.0 - 168.5 rn /m in Ch2,114.3 - 632.3 p /m in Ch3, and -323.2 - -894.7 pm /m in Ch4. Conclusion. A comparative statistical analysis using paired sample t-test about GroupⅠ Vs GroupⅡ under distal cantilevered load shows that there are statistical significant differences for all 4 channels (P<0.05). | en |
dc.language.iso | en | en |
dc.publisher | 대한치과보철학회 | en |
dc.subject | Strain gauge analysis | en |
dc.subject | Implant-supported cantilevered fixed prosthesis | en |
dc.subject | Digital push-pull gauge | en |
dc.title | A strain gauge analysis of implant-supported cantilevered fixed prosthesis under distal static load | en |
dc.type | Article | en |
dc.contributor.AlternativeAuthor | 손병섭 | - |
dc.contributor.AlternativeAuthor | 허성주 | - |
dc.contributor.AlternativeAuthor | 장익태 | - |
dc.contributor.AlternativeAuthor | 곽재영 | - |
dc.contributor.AlternativeAuthor | 김성균 | - |
- Appears in Collections:
- Files in This Item:
Item View & Download Count
Items in S-Space are protected by copyright, with all rights reserved, unless otherwise indicated.