Performance Evaluation of the GlucoDr Plus Glucometer
Cited 2 time in
Web of Science
Cited 2 time in Scopus
- Authors
- Issue Date
- 2010-04
- Publisher
- MARY ANN LIEBERT INC
- Citation
- DIABETES TECHNOLOGY & THERAPEUTICS; Vol.12 4; 307-312
- Abstract
- Background: Because strict glucose control is important for reducing the complications of diabetes, the self-monitoring of blood glucose is one of the fundamental treatment modalities. Many glucometers have been developed. In the present study, we evaluated a new glucometer: GlucoDr (TM) Plus (Allmedicus, Anyang, Gyeonggi-do, Republic of Korea). Methods: The evaluation was performed based on Clinical and Laboratory Standards Institute guidelines. Interferences by ascorbic acid, uric acid, maltose, and acetaminophen were examined, and the performance of the unit was compared to those of six other glucometers. The effects of hematocrit, of oxygen partial pressure (PaO(2)), and of multiple users were also evaluated. Results: Within-run, between-run, between- day, and total imprecision (coefficients of variation) were 0.99-4.98%. Satisfactory linearity was found for glucose concentrations of 32.5-786.5 mg/dL (R(2) = 0.9985). A comparison with the reference laboratory method showed close concordance over the entire range of concentrations evaluated (R(2) = 0.9869). No significant effects were noted due to added interferents, hematocrit, and PaO(2). Conclusions: The GlucoDr Plus showed acceptable performance in terms of precision and linearity. It was minimally affected by various interferents. GlucoDr Plus is suitable for the self-monitoring of blood glucose by patients with diabetes.
- ISSN
- 1520-9156
- Language
- English
- Files in This Item:
Item View & Download Count
Items in S-Space are protected by copyright, with all rights reserved, unless otherwise indicated.