Publications
Detailed Information
Fletcher-Suit Applicator가 쎄시움-137 선량분포에 미치는 영향에 관한 측정 : The Measurements of Dosimetry of Cs -137 Sources with the Fletcher-Suit Applicator
Cited 0 time in
Web of Science
Cited 0 time in Scopus
- Authors
- Issue Date
- 1981-06
- Publisher
- 서울대학교 의과대학
- Citation
- Seoul J Med, Vol.22 No.2, pp. 289-292
- Abstract
- Fletcher'Suit applicator was designed with specific
purpose for the treatment of carcinoma of the uterine
cervix and has been widely used for the intracavitary
radiation.
In many instances. calculation of dosage distributions
about intracavitary applicator doses not require
corrections for radiation by applicator.
However. in the case of the Fletcher-Suit applicator
loaded with cesium in place of radium tubes. correction
should be made for absorption by the applicator.
So the measurements of transmission ratios has
made in polystyrene phantom using PC•12 Computer
densitometer.
The results show that there is a 10%~30% reduction
dose to the region of the hladder trigone and
anterior rectum