Publications

Detailed Information

Fletcher-Suit Applicator가 쎄시움-137 선량분포에 미치는 영향에 관한 측정 : The Measurements of Dosimetry of Cs -137 Sources with the Fletcher-Suit Applicator

Cited 0 time in Web of Science Cited 0 time in Scopus
Authors

박찬일; 하성환

Issue Date
1981-06
Publisher
서울대학교 의과대학
Citation
Seoul J Med, Vol.22 No.2, pp. 289-292
Abstract
Fletcher'Suit applicator was designed with specific
purpose for the treatment of carcinoma of the uterine
cervix and has been widely used for the intracavitary
radiation.
In many instances. calculation of dosage distributions
about intracavitary applicator doses not require
corrections for radiation by applicator.
However. in the case of the Fletcher-Suit applicator
loaded with cesium in place of radium tubes. correction
should be made for absorption by the applicator.
So the measurements of transmission ratios has
made in polystyrene phantom using PC•12 Computer
densitometer.
The results show that there is a 10%~30% reduction
dose to the region of the hladder trigone and
anterior rectum
ISSN
0582-6802
Language
Korean
URI
https://hdl.handle.net/10371/9504
Files in This Item:
Appears in Collections:

Altmetrics

Item View & Download Count

  • mendeley

Items in S-Space are protected by copyright, with all rights reserved, unless otherwise indicated.