Publications
Detailed Information
Gain Switching for a Detection System to Accommodate a Newly Developed MALDI-Based Quantification Method
DC Field | Value | Language |
---|---|---|
dc.contributor.author | Ahn, Sung Hee | - |
dc.contributor.author | Hyeon, Taeghwan | - |
dc.contributor.author | Kim, Myung Soo | - |
dc.contributor.author | Moon, Jeong Hee | - |
dc.date.accessioned | 2020-04-27T13:29:38Z | - |
dc.date.available | 2020-04-27T13:29:38Z | - |
dc.date.created | 2018-09-10 | - |
dc.date.issued | 2017-09 | - |
dc.identifier.citation | Journal of the American Society for Mass Spectrometry, Vol.28 No.9, pp.1987-1990 | - |
dc.identifier.issn | 1044-0305 | - |
dc.identifier.other | 52186 | - |
dc.identifier.uri | https://hdl.handle.net/10371/165914 | - |
dc.description.abstract | In matrix-assisted laser desorption ionization time-of-flight mass spectrometry (MALDI-TOF), matrix-derived ions are routinely deflected away to avoid problems with ion detection. This, however, limits the use of a quantification method that utilizes the analyte-to-matrix ion abundance ratio. In this work, we will show that it is possible to measure this ratio by a minor instrumental modification of a simple form of MALDI-TOF. This involves detector gain switching. | - |
dc.language | 영어 | - |
dc.publisher | Elsevier BV | - |
dc.title | Gain Switching for a Detection System to Accommodate a Newly Developed MALDI-Based Quantification Method | - |
dc.type | Article | - |
dc.contributor.AlternativeAuthor | 김명수 | - |
dc.contributor.AlternativeAuthor | 현택환 | - |
dc.identifier.doi | 10.1007/s13361-017-1711-2 | - |
dc.citation.journaltitle | Journal of the American Society for Mass Spectrometry | - |
dc.identifier.wosid | 000407776200027 | - |
dc.identifier.scopusid | 2-s2.0-85027414267 | - |
dc.citation.endpage | 1990 | - |
dc.citation.number | 9 | - |
dc.citation.startpage | 1987 | - |
dc.citation.volume | 28 | - |
dc.identifier.sci | 000407776200027 | - |
dc.description.isOpenAccess | N | - |
dc.contributor.affiliatedAuthor | Hyeon, Taeghwan | - |
dc.contributor.affiliatedAuthor | Kim, Myung Soo | - |
dc.type.docType | Article | - |
dc.description.journalClass | 1 | - |
dc.subject.keywordPlus | LASER-DESORPTION IONIZATION | - |
dc.subject.keywordPlus | MASS-SPECTROMETRY | - |
dc.subject.keywordPlus | PEPTIDES | - |
dc.subject.keywordPlus | DESORPTION/IONIZATION | - |
dc.subject.keywordAuthor | Detector gain switching | - |
dc.subject.keywordAuthor | MALDI-TOF instrument | - |
dc.subject.keywordAuthor | MALDI quantification | - |
- Appears in Collections:
- Files in This Item:
- There are no files associated with this item.
Item View & Download Count
Items in S-Space are protected by copyright, with all rights reserved, unless otherwise indicated.