Publications

Detailed Information

Three-dimensional analysis of electrode behavior in a human cochlear model

Cited 0 time in Web of Science Cited 0 time in Scopus
Authors

Lim, Yoon Seob; Park, Se Ik; Kim, Sung June

Issue Date
2005-10
Publisher
Elsevier
Citation
Med Eng Phys 2005;27(8):695-703
Keywords
Cochlear implantCochlear electrodeCochlear electrode
Abstract
Three-dimensional (3D) finite element analysis is used in this study to model the mechanical effects of the electrode in a cochlear implant.
We employ six electrodes with different stiffness produced by different arrangements of metal wires. Different wire arrangements are generated
by changing the fitness function of a genetic program. The human cochlea is modeled by the spiral-approximation method. Reconstructed
three-dimensional CT images are used to model the real insertion condition. The contact pressure at the tip and the insertion force are found
to be highest when the wires are stacked horizontally. Axial rotation of the electrode has minimal effect on the stimulating current spread.
The electrode does not contact the basilar membrane. The results indicate that considering the electrode stiffness is important to minimizing
mechanical trauma in cochlear implantation.
ISSN
1350-4533
Language
English
URI
https://hdl.handle.net/10371/8858
Files in This Item:
There are no files associated with this item.
Appears in Collections:

Altmetrics

Item View & Download Count

  • mendeley

Items in S-Space are protected by copyright, with all rights reserved, unless otherwise indicated.

Share