Three-dimensional analysis of electrode behavior in a human cochlear model
Cited 0 time in
Web of Science
Cited 0 time in Scopus
- Authors
- Issue Date
- 2005-10
- Publisher
- Elsevier
- Citation
- Med Eng Phys 2005;27(8):695-703
- Keywords
- Cochlear implant ; Cochlear electrode ; Cochlear electrode
- Abstract
- Three-dimensional (3D) finite element analysis is used in this study to model the mechanical effects of the electrode in a cochlear implant.
We employ six electrodes with different stiffness produced by different arrangements of metal wires. Different wire arrangements are generated
by changing the fitness function of a genetic program. The human cochlea is modeled by the spiral-approximation method. Reconstructed
three-dimensional CT images are used to model the real insertion condition. The contact pressure at the tip and the insertion force are found
to be highest when the wires are stacked horizontally. Axial rotation of the electrode has minimal effect on the stimulating current spread.
The electrode does not contact the basilar membrane. The results indicate that considering the electrode stiffness is important to minimizing
mechanical trauma in cochlear implantation.